A System for Stereotactic Radiosurgery with a Linear Accelerator

Overview
Specialties Oncology
Radiology
Date 1988 Feb 1
PMID 3276655
Citations 147
Authors
Affiliations
Soon will be listed here.
Abstract

A small field irradiation technique to deliver high doses of single fraction photon radiation to small, precisely located volumes (0.5 to 8 cm3) within the brain has been developed. Our method uses a modified Brown-Roberts-Wells (BRS), CT-guided, stereotactic system and a 6 MV linear accelerator equipped with a special collimator (diameters of 12.5 mm to 30.0 mm projected to isocenter) located 23 cm from isocenter. Target localization via planar angiography has been added. Treatment consists of a series of arcing beams using both gantry and couch rotations. During treatment, the patient's head is immobilized independently of the radiotherapy couch and is precisely positioned without reference to room lasers or light field. A precise verification of alignment precedes each treatment. Extensive performance tests have shown that a target, localized by CT, can be irradiated with a positional accuracy of 2.4 mm in any direction with 95% confidence. If angiography is used for localization, the results are better. The dose 1.0 cm outside the target volume is less than 20% of the prescribed dose for a medium sized collimator.

Citing Articles

Advancements and challenges in brain cancer therapeutics.

Bai F, Deng Y, Li L, Lv M, Razzokov J, Xu Q Exploration (Beijing). 2024; 4(6):20230177.

PMID: 39713205 PMC: 11655316. DOI: 10.1002/EXP.20230177.


Development of dedicated phantom for end-to-end testing of multiple image-guidance configurations incorporating light-section-based surface guidance system.

Hayashi N, Saito T, Ouchi K, Amma H, Muraki Y, Yasui K J Appl Clin Med Phys. 2024; 26(1):e14517.

PMID: 39600111 PMC: 11712516. DOI: 10.1002/acm2.14517.


Quantification of uncertainties in reference and relative dose measurements, dose calculations, and patient setup in modern external beam radiotherapy.

Kinoshita N, Shimizu M, Motegi K, Tsuruta Y, Takakura T, Oguchi H Radiol Phys Technol. 2024; 18(1):58-77.

PMID: 39541009 PMC: 11876197. DOI: 10.1007/s12194-024-00856-0.


Historical Review of Stereotactic Radiosurgery in Juntendo University.

Naoi Y Juntendo Iji Zasshi. 2024; 68(5):459-464.

PMID: 39081587 PMC: 11284295. DOI: 10.14789/jmj.JMJ22-0029-R.


Technical Notes: Robustness of three-dimensional treatment and imaging isocenter testing using a new gel dosimeter and kilovoltage CBCT.

Oshika R, Tachibana H, Seki K, Tachibana R, Moriya S, Sakae T J Appl Clin Med Phys. 2024; 25(9):e14439.

PMID: 39031633 PMC: 11492395. DOI: 10.1002/acm2.14439.