» Articles » PMID: 36143066

Evaluation of a New Echocardiographic Tool for Cardiac Output Monitoring: An Experimental Study on A Controlled Hemorrhagic Shock Model in Anesthetized Piglets

Overview
Journal J Clin Med
Specialty General Medicine
Date 2022 Sep 23
PMID 36143066
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Cardiac output (CO) monitoring is recommended in patients with shock. The search for a reliable, rapid, and noninvasive tool is necessary for clinical practice. A new echocardiographic CO flow index (COF) is the automatic calculation of the sub-aortic VTI multiplied by the automatic calculation of the heart rate (HR). The primary objective of this study was to show the correlation between COF and CO measured by thermodilution (COth) in a controlled hemorrhagic shock model in anesthetized piglets. Secondary objectives were to show the correlation between COth and CO calculated from left outflow tract (LVOT) measurement and manual VTI (COman), and CO measured by LVOT measurement and VTIauto (COauto).

Methods: Prospective interventional experimental study. In seventeen ventilated and anesthetized piglets, a state of hemorrhagic shock was induced, maintained, then resuscitated and stabilized. The gold standard for CO and stroke volume measurement was thermodilution (COth).

Results: 191 measurements were performed. The correlation coefficients (r) between COth and COF, COman, and COauto were 0.73 [0.62; 0.81], 0.66 [0.56; 0.74], and 0.73 [0.63; 0.81], respectively.

Conclusions: In this study, the COF appears to have a strong correlation to the COth. This automatic index, which takes into account the HR and does not require the measurement of LVOT, could be a rapidly obtained index in clinical practice.

Citing Articles

Association between speckle tracking echocardiography and pressure-volume loops during cardiogenic shock development.

Frederiksen P, Linde L, Gregers E, Udesen N, Helgestad O, Banke A Open Heart. 2024; 11(1).

PMID: 38782543 PMC: 11116883. DOI: 10.1136/openhrt-2023-002512.

References
1.
Koobi T, Kaukinen S, Turjanmaa V . Cardiac output can be reliably measured noninvasively after coronary artery bypass grafting operation. Crit Care Med. 1999; 27(10):2206-11. DOI: 10.1097/00003246-199910000-00023. View

2.
Temporelli P, Scapellato F, Eleuteri E, Imparato A, Giannuzzi P . Doppler echocardiography in advanced systolic heart failure: a noninvasive alternative to Swan-Ganz catheter. Circ Heart Fail. 2010; 3(3):387-94. DOI: 10.1161/CIRCHEARTFAILURE.108.809590. View

3.
Zieleskiewicz L, Claret P, Muller L, de La Coussaye J, Lefrant J, Schuster I . Global longitudinal strain changes during hemorrhagic shock: An experimental study. Turk J Emerg Med. 2020; 20(3):97-104. PMC: 7416855. DOI: 10.4103/2452-2473.290066. View

4.
Yamada T, Tsutsui M, Sugo Y, Sato T, Akazawa T, Sato N . Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output. Anesth Analg. 2012; 115(1):82-7. DOI: 10.1213/ANE.0b013e31824e2b6c. View

5.
Cecconi M, De Backer D, Antonelli M, Beale R, Bakker J, Hofer C . Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med. 2014; 40(12):1795-815. PMC: 4239778. DOI: 10.1007/s00134-014-3525-z. View