A Multi-modality Physical Phantom for Mimicking Tumor Heterogeneity Patterns in PET/CT and PET/MRI

Overview
Journal Med Phys
Specialty Biophysics
Date 2022 Jul 15
PMID 35838056
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Hybrid imaging (e.g., positron emission tomography [PET]/computed tomography [CT], PET/magnetic resonance imaging [MRI]) helps one to visualize and quantify morphological and physiological tumor characteristics in a single study. The noninvasive characterization of tumor heterogeneity is essential for grading, treatment planning, and following-up oncological patients. However, conventional (CONV) image-based parameters, such as tumor diameter, tumor volume, and radiotracer activity uptake, are insufficient to describe tumor heterogeneities. Here, radiomics shows promise for a better characterization of tumors. Nevertheless, the validation of such methods demands imaging objects capable of reflecting heterogeneities in multi-modality imaging. We propose a phantom to simulate tumor heterogeneity repeatably in PET, CT, and MRI.

Methods: The phantom consists of three 50-ml plastic tubes filled partially with acrylic spheres of S1: 1.6 mm, S2: 50%(1.6 mm)/50%(6.3 mm), or S3: 6.3-mm diameter. The spheres were fixed to the bottom of each tube by a plastic grid, yielding one sphere free homogeneous region and one heterogeneous (S1, S2, or S3) region per tube. A 3-tube phantom and its replica were filled with a fluorodeoxyglucose (18F) solution for test-retest measurements in a PET/CT Siemens TPTV and a PET/MR Siemens Biograph mMR system. A number of 42 radiomic features (10 first order and 32 texture features) were calculated for each phantom region and imaging modality. Radiomic features stability was evaluated through coefficients of variation (COV) across phantoms and scans for PET, CT, and MRI. Further, the Wilcoxon test was used to assess the capability of stable features to discriminate the simulated phantom regions.

Results: The different patterns (S1-S3) did present visible heterogeneity in all imaging modalities. However, only for CT and MRI, a clear visual difference was present between the different patterns. Across all phantom regions in PET, CT, and MR images, 10, 16, and 21 features out of 42 evaluated features in total had a COV of 10% or less. In particular, CONV, histogram, and gray-level run length matrix features showed high repeatability for all the phantom regions and imaging modalities. Several of repeatable texture features allowed the image-based discrimination of the different phantom regions (p < 0.05). However, depending on the feature, different pattern discrimination capabilities were found for the different imaging modalities.

Conclusion: The proposed phantom appears suitable for simulating heterogeneities in PET, CT, and MRI. We demonstrate that it is possible to select radiomic features for the readout of the phantom. Most of these features had been shown to be relevant in previous clinical studies.

Citing Articles

Multicentric study on the reproducibility and robustness of PET-based radiomics features with a realistic activity painting phantom.

Kallos-Balogh P, Vas N, Toth Z, Szakall S, Szabo P, Garai I PLoS One. 2024; 19(10):e0309540.

PMID: 39446842 PMC: 11500893. DOI: 10.1371/journal.pone.0309540.


A customized anthropomorphic 3D-printed phantom to reproducibility assessment in computed tomography: an oncological case study.

Cavaliere C, Baldi D, Brancato V, Aiello M, Salvatore M Front Oncol. 2023; 13:1123796.

PMID: 37700836 PMC: 10493384. DOI: 10.3389/fonc.2023.1123796.


A multi-modality physical phantom for mimicking tumor heterogeneity patterns in PET/CT and PET/MRI.

Valladares A, Beyer T, Papp L, Salomon E, Rausch I Med Phys. 2022; 49(9):5819-5829.

PMID: 35838056 PMC: 9543355. DOI: 10.1002/mp.15853.

References
1.
Harry V, Semple S, Gilbert F, Parkin D . Diffusion-weighted magnetic resonance imaging in the early detection of response to chemoradiation in cervical cancer. Gynecol Oncol. 2008; 111(2):213-20. DOI: 10.1016/j.ygyno.2008.07.048. View

2.
Valladares A, Beyer T, Papp L, Salomon E, Rausch I . A multi-modality physical phantom for mimicking tumor heterogeneity patterns in PET/CT and PET/MRI. Med Phys. 2022; 49(9):5819-5829. PMC: 9543355. DOI: 10.1002/mp.15853. View

3.
Liu Z, Wang S, Dong D, Wei J, Fang C, Zhou X . The Applications of Radiomics in Precision Diagnosis and Treatment of Oncology: Opportunities and Challenges. Theranostics. 2019; 9(5):1303-1322. PMC: 6401507. DOI: 10.7150/thno.30309. View

4.
Yip S, Aerts H . Applications and limitations of radiomics. Phys Med Biol. 2016; 61(13):R150-66. PMC: 4927328. DOI: 10.1088/0031-9155/61/13/R150. View

5.
Orlhac F, Soussan M, Chouahnia K, Martinod E, Buvat I . 18F-FDG PET-Derived Textural Indices Reflect Tissue-Specific Uptake Pattern in Non-Small Cell Lung Cancer. PLoS One. 2015; 10(12):e0145063. PMC: 4682929. DOI: 10.1371/journal.pone.0145063. View