» Articles » PMID: 35633488

Performance Evaluation of Three DNA Sample Tracking Tools in a Whole Exome Sequencing Workflow

Overview
Journal Mol Diagn Ther
Date 2022 May 28
PMID 35633488
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Next-generation sequencing applications are becoming indispensable for clinical diagnostics. These experiments require numerous wet- and dry-laboratory steps, each one increasing the probability of a sample swap or contamination. Therefore, identity confirmation at the end of the process is recommended to ensure the right data are used for each patient.

Methods: We tested three commercially available, single nucleotide polymorphism (SNP)-based sample tracking kits in a diagnostic workflow to evaluate their ease of use and performance. The coverage uniformity, on-target specificity, sample identification, and genotyping performance were determined to assess the reliability and cost effectiveness of each kit.

Results And Discussion: Hands-on time and manual steps are almost identical for the kits from pxlence and Nimagen. The Swift kit has an extra purification step, making it the longest and most demanding protocol. Furthermore, the Swift kit failed to correctly genotype 26 of the 46 samples. The Nimagen kit identified all but one sample and the pxlence kit unambiguously identified all samples, making it the most reliable and robust kit of this evaluation. The Nimagen kit showed poor on-target mapping rates, resulting in deeper sequencing needs and higher sequencing costs compared with the other two kits.

Conclusion: Our conclusion is that the Human Sample ID kit from pxlence is the most cost effective of the three tested tools for DNA sample tracking and identification.

Citing Articles

Sample Tracking Tool: A Comprehensive Approach Based on OpenArray Technology and R Scripting for Genomic Sample Monitoring.

Trastulli G, Calvino G, Papasergi B, Megalizzi D, Peconi C, Zampatti S Diagnostics (Basel). 2025; 15(2).

PMID: 39857032 PMC: 11763353. DOI: 10.3390/diagnostics15020149.

References
1.
Kim J, Park W, Kim N, Jang S, Chun S, Sung C . Good Laboratory Standards for Clinical Next-Generation Sequencing Cancer Panel Tests. J Pathol Transl Med. 2017; 51(3):191-204. PMC: 5445206. DOI: 10.4132/jptm.2017.03.14. View

2.
Zhi X, Han Y, Li Y, Wang G, Feng L, Yang B . Root growth and spatial distribution characteristics for seedlings raised in substrate and transplanted cotton. PLoS One. 2017; 12(12):e0190032. PMC: 5741230. DOI: 10.1371/journal.pone.0190032. View