Blood Pressure Continuous Measurement Through a Wearable Device: Development and Validation of a Cuffless Method

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2021 Nov 13
PMID 34770641
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The present study aims to develop and validate a cuffless method for blood pressure continuous measurement through a wearable device. The goal is achieved according to the time-delay method, with the guiding principle of the time relation it takes for a blood volume to travel from the heart to a peripheral site. Inversely proportional to the blood pressure, this time relation is obtained as the time occurring between the R peak of the electrocardiographic signal and a marker point on the photoplethysmographic wave. Such physiological signals are recorded by using L.I.F.E. Italia's wearable device, made of a sensorized shirt and wristband. A linear regression model is implemented to estimate the corresponding blood pressure variations from the obtained time-delay and other features of the photoplethysmographic wave. Then, according to the international standards, the model performance is assessed, comparing the estimates with the measurements provided by a certified digital sphygmomanometer. According to the standards, the results obtained during this study are notable, with 85% of the errors lower than 10 mmHg and a mean absolute error lower than 7 mmHg. In conclusion, this study suggests a time-delay method for continuous blood pressure estimates with good performance, compared with a reference device based on the oscillometric technique.

Citing Articles

Wearable devices for patient monitoring in the intensive care unit.

Angelucci A, Greco M, Cecconi M, Aliverti A Intensive Care Med Exp. 2025; 13(1):26.

PMID: 40016479 PMC: 11868008. DOI: 10.1186/s40635-025-00738-8.


Assessing Usability of Smartwatch Digital Health Devices for Home Blood Pressure Monitoring among Glaucoma Patients.

Bhanvadia S, Brar M, Delavar A, Tavakoli K, Radha Saseendrakumar B, Weinreb R Informatics (MDPI). 2023; 9(4).

PMID: 36873830 PMC: 9980658. DOI: 10.3390/informatics9040079.

References
1.
Sarmento A, Vignati C, Paolillo S, Lombardi C, Scoccia A, Nicoli F . Qualitative and quantitative evaluation of a new wearable device for ECG and respiratory Holter monitoring. Int J Cardiol. 2018; 272:231-237. DOI: 10.1016/j.ijcard.2018.06.044. View

2.
Elgendi M, Fletcher R, Liang Y, Howard N, Lovell N, Abbott D . The use of photoplethysmography for assessing hypertension. NPJ Digit Med. 2019; 2:60. PMC: 6594942. DOI: 10.1038/s41746-019-0136-7. View

3.
Contini M, Sarmento A, Gugliandolo P, Leonardi A, Longinotti-Buitoni G, Minella C . Validation of a new wearable device for type 3 sleep test without flowmeter. PLoS One. 2021; 16(4):e0249470. PMC: 8051765. DOI: 10.1371/journal.pone.0249470. View

4.
Garg R, Malhotra V, Kumar A, Dhar U, Tripathi Y . Effect of isometric handgrip exercise training on resting blood pressure in normal healthy adults. J Clin Diagn Res. 2014; 8(9):BC08-10. PMC: 4225874. DOI: 10.7860/JCDR/2014/8908.4850. View

5.
Sharman J, LaGerche A . Exercise blood pressure: clinical relevance and correct measurement. J Hum Hypertens. 2014; 29(6):351-8. DOI: 10.1038/jhh.2014.84. View