Comparison of Accuracy of Two Uncalibrated Pulse Contour Cardiac Output Monitors in Off-pump Coronary Artery Bypass Surgery Patients Using Pulmonary Artery Catheter-thermodilution As a Reference

Overview
Journal BMC Anesthesiol
Publisher Biomed Central
Specialty Anesthesiology
Date 2021 Jul 11
PMID 34246222
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Cardiac output (CO) is a key measure of adequacy of organ and tissue perfusion, especially in critically ill or complex surgical patients. CO monitoring technology continues to evolve. Recently developed CO monitors rely on unique algorithms based on pulse contour analysis of an arterial blood pressure (ABP) waveform. The objective of this investigation was to compare the accuracy of two monitors using different methods of pulse contour analysis - the Retia Argos device and the Edwards Vigileo-FloTrac device - with pulmonary artery catheter (PAC)-thermodilution as a reference.

Methods: Fifty-eight patients undergoing off-pump coronary artery bypass surgery formed the study cohort. A total of 572 triplets of CO measurements from each device - Argos, Vigileo-FloTrac (third generation), and thermodilution - were available before and after interventions (e.g., vasopressors, fluids, and inotropes). Bland-Altman analysis accounting for repeated measurements per subject and concordance analysis were applied to assess the accuracy of the CO values and intervention-induced CO changes of each pulse contour device against thermodilution. Cluster bootstrapping was employed to statistically compare the root-mean-squared-errors (RMSE = √(μ + σ), where μ and σ are the Bland-Altman bias and precision errors) and concordance rates of the two devices.

Results: The RMSE (mean (95% confidence intervals)) for CO values was 1.16 (1.00-1.32) L/min for the Argos device and 1.54 (1.33-1.77) L/min for the Vigileo-FloTrac device; the concordance rate for intervention-induced CO changes was 87 (82-92)% for the Argos device and 72 (65-78)% for the Vigileo-FloTrac device; and the RMSE for the CO changes was 17 (15-19)% for the Argos device and 21 (19-23)% for the Vigileo-FloTrac device (p < 0.0167 for all comparisons).

Conclusions: In comparison with CO measured by the PAC, the Argos device proved to be more accurate than the Vigileo-FloTrac device in CO trending and absolute CO measurement in patients undergoing off-pump coronary artery bypass surgery.

Citing Articles

Occurrence of Low Cardiac Index During Normotensive Periods in Cardiac Surgery: A Prospective Cohort Study Using Continuous Noninvasive Cardiac Output Monitoring.

Goeddel L, Koffman L, Hernandez M, Whitman G, Parikh C, Lima J Anesth Analg. 2024; 140(1):77-86.

PMID: 39207928 PMC: 11649474. DOI: 10.1213/ANE.0000000000007206.


Bias, trending ability and diagnostic performance of a non-calibrated multi-beat analysis continuous cardiac output monitor to identify fluid responsiveness in critically ill patients.

Bitker L, Noirot I, Chauvelot L, Mezidi M, Dhelft F, Gaillet M Crit Care Resusc. 2024; 26(2):108-115.

PMID: 39072233 PMC: 11282375. DOI: 10.1016/j.ccrj.2024.04.003.


Pulmonary artery catheter receipt among cardiac surgery patients from the national inpatient sample (1999-2019): Prevalence, predictors and hospitalization charges.

Beydoun H, Beydoun M, Eid S, Zonderman A Heliyon. 2024; 10(3):e24902.

PMID: 38317919 PMC: 10839978. DOI: 10.1016/j.heliyon.2024.e24902.


Anesthetic management of a huge retroperitoneal leiomyoma: a case report.

Shi Y, Zhu B, Zhang Y, Huang Y Perioper Med (Lond). 2023; 12(1):64.

PMID: 38017529 PMC: 10683212. DOI: 10.1186/s13741-023-00352-w.


Tissue perfusion pressure enables continuous hemodynamic evaluation and risk prediction in the intensive care unit.

Chandrasekhar A, Padros-Valls R, Pallares-Lopez R, Palanques-Tost E, Houstis N, Sundt T Nat Med. 2023; 29(8):1998-2006.

PMID: 37550417 DOI: 10.1038/s41591-023-02474-6.


References
1.
Bland J, Altman D . Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007; 17(4):571-82. DOI: 10.1080/10543400701329422. View

2.
Saugel B, Heeschen J, Hapfelmeier A, Romagnoli S, Greiwe G . Cardiac output estimation using multi-beat analysis of the radial arterial blood pressure waveform: a method comparison study in patients having off-pump coronary artery bypass surgery using intermittent pulmonary artery thermodilution as the.... J Clin Monit Comput. 2019; 34(4):649-654. PMC: 7367911. DOI: 10.1007/s10877-019-00375-z. View

3.
Critchley L, Lee A, Ho A . A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010; 111(5):1180-92. DOI: 10.1213/ANE.0b013e3181f08a5b. View

4.
Suehiro K, Tanaka K, Mikawa M, Uchihara Y, Matsuyama T, Matsuura T . Improved Performance of the Fourth-Generation FloTrac/Vigileo System for Tracking Cardiac Output Changes. J Cardiothorac Vasc Anesth. 2014; 29(3):656-62. DOI: 10.1053/j.jvca.2014.07.022. View

5.
Hao G, Liu Y, Ma G, Hou J, Zhu D, Liu L . Reliability of three-dimensional color flow Doppler and two-dimensional pulse wave Doppler transthoracic echocardiography for estimating cardiac output after cardiac surgery. Cardiovasc Ultrasound. 2019; 17(1):5. PMC: 6448273. DOI: 10.1186/s12947-019-0155-1. View