Diagnostic Accuracy of LAMP Versus PCR over the Course of SARS-CoV-2 Infection

Overview
Publisher Elsevier
Date 2021 Apr 16
PMID 33862213
Citations 40
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Reverse transcription loop-mediated isothermal amplification (RT-LAMP) has been validated to diagnose several viral infections. However, its diagnostic accuracy in detecting SARS-CoV-2 in real-life clinical settings remains unclear. This study aimed to determine the diagnostic sensitivity and specificity of RT-LAMP compared to reverse transcription-quantitative polymerase chain reaction (RT-qPCR) over the disease course of COVID-19.

Methods: A total of 124 nasopharyngeal swab samples obtained from 24 COVID-19 patients were tested by RT-LAMP and RT-qPCR. Sensitivities and specificities of RT-LAMP compared with RT-qPCR were analyzed as a function of time from onset.

Results: Up to the 9th day after onset, the RT-LAMP had a positivity of 92.8%, and the sensitivity and specificity compared with RT-qPCR was 100%. However, after the 10th day after onset, the positivity of RT-LAMP decreased to less than 25%, and the concordance of positivity between the two methods was below 60%. The limit of detection of RT-LAMP was 6.7 copies/reaction.

Conclusions: Until the 9th day after the onset of symptoms, RT-LAMP had the same diagnostic accuracy as RT-qPCR. These findings suggest that RT-LAMP can be used as a diagnostic tool for COVID-19 as an alternative to RT-qPCR in the acute symptomatic phase of COVID-19.

Citing Articles

Optimization and Benchmarking of RT-LAMP-CRISPR-Cas12a for the Detection of SARS-CoV-2 in Saliva.

Lynch C, Drummond R, Jelley L, Baker L, Smit E, Fleming R Int J Mol Sci. 2025; 26(5).

PMID: 40076433 PMC: 11899638. DOI: 10.3390/ijms26051806.


RUNCOV: a one-pot triplex real-time RT-LAMP as a point-of-care diagnostic tool for detecting SARS-CoV-2.

Robene I, Jouen E, Maillot-Lebon V, Fenelon B, Hascoat J, Pecrix Y Biol Methods Protoc. 2025; 10(1):bpaf010.

PMID: 40046730 PMC: 11882304. DOI: 10.1093/biomethods/bpaf010.


Evaluation of three rapid assays for detecting Mycoplasma pneumoniae in nasopharyngeal specimens.

Yang C, Wang Z, Kong L, Du J, Yi J AMB Express. 2024; 14(1):134.

PMID: 39694968 PMC: 11655899. DOI: 10.1186/s13568-024-01782-5.


Commercial Opportunity or Addressing Unmet Needs-Loop-Mediated Isothermal Amplification (LAMP) as the Future of Rapid Diagnostic Testing?.

Feddema J, Fernald K, Keijser B, Kieboom J, van de Burgwal L Diagnostics (Basel). 2024; 14(17).

PMID: 39272630 PMC: 11394392. DOI: 10.3390/diagnostics14171845.


Cross comparison of alternative diagnostic protocols including substitution to the clinical sample, RNA extraction method and nucleic acid amplification technology for COVID-19 diagnosis.

Segura-Ulate I, Apu N, Cortes B, Querol-Audi J, Zaldivar Y, Ortega C Front Mol Biosci. 2024; 11:1445142.

PMID: 39247206 PMC: 11377848. DOI: 10.3389/fmolb.2024.1445142.


References
1.
Augustine R, Hasan A, Das S, Ahmed R, Mori Y, Notomi T . Loop-Mediated Isothermal Amplification (LAMP): A Rapid, Sensitive, Specific, and Cost-Effective Point-of-Care Test for Coronaviruses in the Context of COVID-19 Pandemic. Biology (Basel). 2020; 9(8). PMC: 7464797. DOI: 10.3390/biology9080182. View

2.
Mallett S, Allen A, Graziadio S, Taylor S, Sakai N, Green K . At what times during infection is SARS-CoV-2 detectable and no longer detectable using RT-PCR-based tests? A systematic review of individual participant data. BMC Med. 2020; 18(1):346. PMC: 7609379. DOI: 10.1186/s12916-020-01810-8. View

3.
Sterne J, Murthy S, Diaz J, Slutsky A, Villar J, Angus D . Association Between Administration of Systemic Corticosteroids and Mortality Among Critically Ill Patients With COVID-19: A Meta-analysis. JAMA. 2020; 324(13):1330-1341. PMC: 7489434. DOI: 10.1001/jama.2020.17023. View

4.
Higashimoto Y, Ihira M, Kawamura Y, Inaba M, Shirato K, Suzuki T . Dry loop-mediated isothermal amplification assay for detection of SARS-CoV-2 from clinical specimens. Fujita Med J. 2023; 9(2):84-89. PMC: 10206895. DOI: 10.20407/fmj.2022-003. View

5.
Yan C, Cui J, Huang L, DU B, Chen L, Xue G . Rapid and visual detection of 2019 novel coronavirus (SARS-CoV-2) by a reverse transcription loop-mediated isothermal amplification assay. Clin Microbiol Infect. 2020; 26(6):773-779. PMC: 7144850. DOI: 10.1016/j.cmi.2020.04.001. View