» Articles » PMID: 33525488

A Compact Multi-Distance DCS and Time Domain NIRS Hybrid System for Hemodynamic and Metabolic Measurements

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2021 Feb 2
PMID 33525488
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

In this work, we present a new multi-distance diffuse correlation spectroscopy (DCS) device integrated with a compact state-of-the-art time domain near infrared spectroscopy (TD-NIRS) device. The hybrid DCS and TD-NIRS system allows to retrieve information on blood flow, tissue oxygenation, and oxygen metabolic rate. The DCS device performances were estimated in terms of stability, repeatability, ability in retrieving variations of diffusion coefficient, influence of the tissue optical properties, effect of varying count rates and depth sensitivity. Crosstalk between DCS and TD-NIRS optical signals was also evaluated. Finally, in vivo experiments (venous and arterial cuff occlusions on the arm) were conducted to test the ability of the hybrid system in measuring blood flow variations.

Citing Articles

Fast Multi-Distance Time-Domain NIRS and DCS System for Clinical Applications.

Nabacino M, Amendola C, Contini D, Re R, Spinelli L, Torricelli A Sensors (Basel). 2024; 24(22).

PMID: 39599151 PMC: 11598675. DOI: 10.3390/s24227375.


A personalized clinical assessment: multi-sensor approach for understanding musculoskeletal health in the frail population.

Re R, Scano A, Amata O, Spinelli L, Tomba A, Brambilla C Biomed Eng Online. 2024; 23(1):91.

PMID: 39252062 PMC: 11382456. DOI: 10.1186/s12938-024-01287-z.


Assessment of power spectral density of microvascular hemodynamics in skeletal muscles at very low and low-frequency via near-infrared diffuse optical spectroscopies.

Amendola C, Buttafava M, Carteano T, Contini L, Cortese L, Durduran T Biomed Opt Express. 2023; 14(11):5994-6015.

PMID: 38021143 PMC: 10659778. DOI: 10.1364/BOE.502618.


Advanced Neuromonitoring Modalities on the Horizon: Detection and Management of Acute Brain Injury in Children.

Ko T, Catennacio E, Shin S, Stern J, Massey S, Kilbaugh T Neurocrit Care. 2023; 38(3):791-811.

PMID: 36949362 PMC: 10241718. DOI: 10.1007/s12028-023-01690-9.


A Contact-Sensitive Probe for Biomedical Optics.

Renna M, Peruch A, Sunwoo J, Starkweather Z, Martin A, Franceschini M Sensors (Basel). 2022; 22(6).

PMID: 35336531 PMC: 8953277. DOI: 10.3390/s22062361.


References
1.
Carp S, Farzam P, Redes N, Hueber D, Franceschini M . Combined multi-distance frequency domain and diffuse correlation spectroscopy system with simultaneous data acquisition and real-time analysis. Biomed Opt Express. 2017; 8(9):3993-4006. PMC: 5611918. DOI: 10.1364/BOE.8.003993. View

2.
De Carli A, Andresen B, Giovannella M, Durduran T, Contini D, Spinelli L . Cerebral oxygenation and blood flow in term infants during postnatal transition: BabyLux project. Arch Dis Child Fetal Neonatal Ed. 2019; 104(6):F648-F653. DOI: 10.1136/archdischild-2018-316400. View

3.
Torricelli A, Contini D, Pifferi A, Caffini M, Re R, Zucchelli L . Time domain functional NIRS imaging for human brain mapping. Neuroimage. 2013; 85 Pt 1:28-50. DOI: 10.1016/j.neuroimage.2013.05.106. View

4.
Re R, Pirovano I, Contini D, Spinelli L, Torricelli A . Time Domain Near Infrared Spectroscopy Device for Monitoring Muscle Oxidative Metabolism: Custom Probe and In Vivo Applications. Sensors (Basel). 2018; 18(1). PMC: 5795927. DOI: 10.3390/s18010264. View

5.
Khalid M, Milej D, Rajaram A, Abdalmalak A, Morrison L, Diop M . Development of a stand-alone DCS system for monitoring absolute cerebral blood flow. Biomed Opt Express. 2019; 10(9):4607-4620. PMC: 6757462. DOI: 10.1364/BOE.10.004607. View