Real-time, Non-invasive Thrombus Detection in an Extracorporeal Circuit Using Micro-optical Thrombus Sensors

Overview
Date 2020 Dec 10
PMID 33300399
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Real-time, non-invasive monitoring of thrombus formation in extracorporeal circuits has yet to be achieved. To address the challenges of conventional optical thrombus detection methods requiring large devices that limit detection capacity, we developed a micro-optical thrombus sensor.

Methods: The proposed micro-optical thrombus sensor can detect the intensity of light scattered by blood at wavelengths of 660 and 855 nm. Two thrombus sensors were installed on in vitro circuit: one at the rotary blood pump and one at a flow channel. To evaluate the variation in the ratio of incident light intensity at each wavelength of the two sensors, (for 660 nm) and (for 855 nm) were defined. Using fresh porcine blood as a working fluid, we performed in vitro tests of haematocrit (Hct) and oxygen saturation (SaO) variation and thrombus detection. Thrombus tests were terminated after or showed a larger change than the maximum range of those in the Hct and SaO variation test.

Results: In all three thrombus detection tests, showed a larger change than the maximum range of those in the Hct and SaO variation test. After the tests, thrombus formation was confirmed in the pump, and there was no thrombus in the flow channel. The results indicate that is an effective parameter for identifying the presence of a thrombus.

Conclusion: Thrombus detection in an extracorporeal circuit using the developed micro-optical sensors was successfully demonstrated in an in vitro test.

Citing Articles

Early thrombus detection in the extracorporeal membrane oxygenation circuit by noninvasive real-time ultrasonic sensors.

Rim G, Hyun K, Cho D, Lim Z, Lee B, Kim K Sci Rep. 2024; 14(1):10438.

PMID: 38714704 PMC: 11076605. DOI: 10.1038/s41598-024-59873-z.


Early Blood Clot Detection Using Forward Scattering Light Measurements Is Not Superior to Delta Pressure Measurements.

Fischbach A, Lamberti M, Simons J, Wrede E, Theissen A, Winnersbach P Biosensors (Basel). 2023; 13(12).

PMID: 38131772 PMC: 10741584. DOI: 10.3390/bios13121012.


Design and Fabrication of a Thin and Micro-Optical Sensor for Rapid Prototyping.

Morita N, Iwasaki W Sensors (Basel). 2023; 23(17).

PMID: 37688114 PMC: 10563074. DOI: 10.3390/s23177658.


Current and future strategies to monitor and manage coagulation in ECMO patients.

Shirejini S, Carberry J, McQuilten Z, Burrell A, Gregory S, Hagemeyer C Thromb J. 2023; 21(1):11.

PMID: 36703184 PMC: 9878987. DOI: 10.1186/s12959-023-00452-z.

References
1.
Sakota D, Murashige T, Kosaka R, Fujiwara T, Nishida M, Maruyama O . Real-Time Observation of Thrombus Growth Process in an Impeller of a Hydrodynamically Levitated Centrifugal Blood Pump by Near-Infrared Hyperspectral Imaging. Artif Organs. 2015; 39(8):714-9. DOI: 10.1111/aor.12560. View

2.
Sakota D, Murashige T, Kosaka R, Nishida M, Maruyama O . Feasibility of the optical imaging of thrombus formation in a rotary blood pump by near-infrared light. Artif Organs. 2014; 38(9):733-40. DOI: 10.1111/aor.12377. View

3.
Li J, Sapkota A, Kikuchi D, Sakota D, Maruyama O, Takei M . Red blood cells aggregability measurement of coagulating blood in extracorporeal circulation system with multiple-frequency electrical impedance spectroscopy. Biosens Bioelectron. 2018; 112:79-85. DOI: 10.1016/j.bios.2018.04.020. View

4.
Fina D, Matteucci M, Jiritano F, Meani P, Lo Coco V, Kowalewski M . Extracorporeal membrane oxygenation without therapeutic anticoagulation in adults: A systematic review of the current literature. Int J Artif Organs. 2020; 43(9):570-578. DOI: 10.1177/0391398820904372. View

5.
Matsuhashi Y, Sameshima K, Yamamoto Y, Umezu M, Iwasaki K . Real-time visualization of thrombus formation at the interface between connectors and tubes in medical devices by using optical coherence tomography. PLoS One. 2017; 12(12):e0188729. PMC: 5720586. DOI: 10.1371/journal.pone.0188729. View