» Articles » PMID: 33260404

Dosimetric Comparison of Intensity-Modulated Radiotherapy, Volumetric Modulated Arc Therapy and Hybrid Three-Dimensional Conformal Radiotherapy/Intensity-Modulated Radiotherapy Techniques for Right Breast Cancer

Overview
Journal J Clin Med
Specialty General Medicine
Date 2020 Dec 2
PMID 33260404
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

This study aimed to compare different types of right breast cancer radiotherapy planning techniques and to estimate the whole-body effective doses and the critical organ absorbed doses. The three planning techniques are intensity-modulated radiotherapy (IMRT), volumetric modulated arc therapy (VMAT; two methods) and hybrid 3D-CRT/IMRT (three-dimensional conformal radiotherapy/intensity-modulated radiotherapy). The VMAT technique includes two methods to deliver a dose: non-continuous partial arc and continuous partial arc. A thermoluminescent dosimeter (TLD) is placed in the RANDO phantom to estimate the organ absorbed dose. Each planning technique applies 50.4 Gy prescription dose and treats critical organs, including the lung and heart. Dose-volume histogram was used to show the planning target volume (V95%), homogeneity index (HI), conformity index (CI), and other optimized indices. The estimation of whole-body effective dose was based on the International Commission on Radiation Protection (ICRP) Publication 60 and 103. The results were as follows: Continuous partial arc and non-continuous partial arc showed the best CI and HI. The heart absorbed doses in the continuous partial arc and hybrid 3D-CRT/IMRT were 0.07 ± 0.01% and 0% (V5% and V10%, respectively). The mean dose of the heart was lowest in hybrid 3D-CRT/IMRT (1.47 Gy ± 0.02). The dose in the left contralateral lung (V5%) was lowest in continuous partial arc (0%). The right ipsilateral lung average dose and V20% are lowest in continuous partial arc. Hybrid 3D-CRT/IMRT has the lowest mean dose to contralateral breast (organs at risk). The whole-body effective doses for ICRP-60 and ICRP-103 were highest in continuous partial arc (2.01 Sv ± 0.23 and 2.89 Sv ± 0.15, respectively). In conclusion, the use of VMAT with continuous arc has a lower risk of radiation pneumonia, while hybrid 3D-CRT/IMRT attain lower secondary malignancy risk and cardiovascular complications.

Citing Articles

Correlations between patient-specific parameters and dosimetric indices for personalized breast cancer radiotherapy.

Costin I, Marcu L Sci Rep. 2024; 14(1):26141.

PMID: 39478060 PMC: 11526019. DOI: 10.1038/s41598-024-75858-4.


A novel x-Ray and γ-Ray combination strategy for radiotherapy after breast-conserving surgery in patients with right breast cancer.

Zhang K, He R, Tang F, Zhou L, Zhang X, Li J Front Oncol. 2024; 14:1397273.

PMID: 39286018 PMC: 11404321. DOI: 10.3389/fonc.2024.1397273.


Evaluation of helical tomotherapy as an alternative for left-sided breast cancer patients not compliant with deep inspiration breath hold.

Kuru O, Ozan Goksel E, Volkan Demircan N, Sengoz M Tech Innov Patient Support Radiat Oncol. 2024; 31:100268.

PMID: 39247163 PMC: 11380380. DOI: 10.1016/j.tipsro.2024.100268.


Dosimetric analysis of six whole-breast irradiation techniques in supine and prone positions.

Kim D, Hong C, Son J, Kim S, Park Y, Chung M Sci Rep. 2024; 14(1):14347.

PMID: 38907042 PMC: 11192744. DOI: 10.1038/s41598-024-65461-y.


Hybrid VMAT-3DCRT as breast cancer treatment improvement tool.

Voyant C, Pinpin M, Leschi D, Prapant S, Savigny F, Acquaviva M Sci Rep. 2024; 13(1):23110.

PMID: 38172237 PMC: 10764879. DOI: 10.1038/s41598-023-50538-x.


References
1.
Mayo C, Urie M, Fitzgerald T . Hybrid IMRT plans--concurrently treating conventional and IMRT beams for improved breast irradiation and reduced planning time. Int J Radiat Oncol Biol Phys. 2005; 61(3):922-32. DOI: 10.1016/j.ijrobp.2004.10.033. View

2.
Lemanski C, Thariat J, Ampil F, Bose S, Vock J, Davis R . Image-guided radiotherapy for cardiac sparing in patients with left-sided breast cancer. Front Oncol. 2014; 4:257. PMC: 4172060. DOI: 10.3389/fonc.2014.00257. View

3.
Sethi R, No H, Jozsef G, Ko J, Formenti S . Comparison of three-dimensional versus intensity-modulated radiotherapy techniques to treat breast and axillary level III and supraclavicular nodes in a prone versus supine position. Radiother Oncol. 2011; 102(1):74-81. DOI: 10.1016/j.radonc.2011.09.008. View

4.
Sakka M, Kunzelmann L, Metzger M, Grabenbauer G . Cardiac dose-sparing effects of deep-inspiration breath-hold in left breast irradiation : Is IMRT more beneficial than VMAT?. Strahlenther Onkol. 2017; 193(10):800-811. DOI: 10.1007/s00066-017-1167-0. View

5.
Vaidya J, Joseph D, Tobias J, Bulsara M, Wenz F, Saunders C . Targeted intraoperative radiotherapy versus whole breast radiotherapy for breast cancer (TARGIT-A trial): an international, prospective, randomised, non-inferiority phase 3 trial. Lancet. 2010; 376(9735):91-102. DOI: 10.1016/S0140-6736(10)60837-9. View