Signal Quality Measure for Pulsatile Physiological Signals Using Morphological Features: Applications in Reliability Measure for Pulse Oximetry

Overview
Date 2020 Sep 1
PMID 32864419
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Pulse oximetry is a noninvasive and low-cost physiological monitor that measures blood oxygen levels. While the noninvasive nature of pulse oximetry is advantageous, the estimates of oxygen saturation generated by these devices are prone to motion artifacts and ambient noise, reducing the reliability of such estimations. Clinicians combat this by assessing the quality of oxygen saturation estimation by visual inspection of the photoplethysmograph (PPG), which represents changes in pulsatile blood volume and is also generated by the pulse oximeter. In this paper, we propose six morphological features that can be used to determine the quality of the PPG signal and generate a signal quality index. Unlike many similar studies, this approach uses machine learning and does not require a separate signal, such as ECG, for reference. Multiple algorithms were tested against 46 30-min PPG segments of patients with cardiovascular and respiratory conditions, including atrial fibrillation, hypoxia, acute heart failure, pneumonia, ARDS, and pulmonary embolism. These signals were independently annotated for signal quality by two clinicians, with the union of their annotations used as the ground-truth. Similar to any physiological signal recorded in a clinical setting, the utilized dataset is also unbalanced in favor of good quality segments. The experiments showed that a cost-sensitive Support Vector Machine (SVM) outperformed other tested methods and was robust to the unbalanced nature of the data. Though the proposed algorithm was tested on PPG signals, the methodology remains agnostic to the dataset used, and may be applied to any type of pulsatile physiological signal.

Citing Articles

Learning under label noise through few-shot human-in-the-loop refinement.

Saeed A, Spathis D, Oh J, Choi E, Etemad A Sci Rep. 2025; 15(1):4276.

PMID: 39905140 PMC: 11794423. DOI: 10.1038/s41598-025-87046-z.


Machine Learning Applied to Reference Signal-Less Detection of Motion Artifacts in Photoplethysmographic Signals: A Review.

Arguello-Prada E, Castillo Garcia J Sensors (Basel). 2024; 24(22).

PMID: 39598970 PMC: 11598458. DOI: 10.3390/s24227193.


Continuous Patient-Independent Estimation of Respiratory Rate and Blood Pressure Using Robust Spectro-Temporal Features Derived From Photoplethysmogram Only.

Sultan M, Saadeh W IEEE Open J Eng Med Biol. 2024; 5:637-649.

PMID: 39184965 PMC: 11342923. DOI: 10.1109/OJEMB.2023.3329728.


Pulse2AI: An Adaptive Framework to Standardize and Process Pulsatile Wearable Sensor Data for Clinical Applications.

Huang S, Jafari R, Mortazavi B IEEE Open J Eng Med Biol. 2024; 5:330-338.

PMID: 38899025 PMC: 11186651. DOI: 10.1109/OJEMB.2024.3398444.


LSTM-based real-time signal quality assessment for blood volume pulse analysis.

Gao H, Zhang C, Pei S, Wu X Biomed Opt Express. 2023; 14(3):1119-1136.

PMID: 36950226 PMC: 10026571. DOI: 10.1364/BOE.477143.


References
1.
Karlen W, Kobayashi K, Ansermino J, Dumont G . Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation. Physiol Meas. 2012; 33(10):1617-29. DOI: 10.1088/0967-3334/33/10/1617. View

2.
Lee C, Yoon H . Medical big data: promise and challenges. Kidney Res Clin Pract. 2017; 36(1):3-11. PMC: 5331970. DOI: 10.23876/j.krcp.2017.36.1.3. View

3.
Orphanidou C, Bonnici T, Charlton P, Clifton D, Vallance D, Tarassenko L . Signal-quality indices for the electrocardiogram and photoplethysmogram: derivation and applications to wireless monitoring. IEEE J Biomed Health Inform. 2014; 19(3):832-8. DOI: 10.1109/JBHI.2014.2338351. View

4.
Li Q, Clifford G . Dynamic time warping and machine learning for signal quality assessment of pulsatile signals. Physiol Meas. 2012; 33(9):1491-501. DOI: 10.1088/0967-3334/33/9/1491. View

5.
Jang D, Kwon U, Yoon S, Park C, Ku Y, Noh S . A Simple and Robust Method for Determining the Quality of Cardiovascular Signals Using the Signal Similarity. Annu Int Conf IEEE Eng Med Biol Soc. 2018; 2018:478-481. DOI: 10.1109/EMBC.2018.8512341. View