In Vitro Simulation of Dental Implant Bridges Removal: Influence of Luting Agent and Abutments Geometry on Retrievability

Overview
Publisher MDPI
Date 2020 Jun 25
PMID 32575862
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Implant fixed dental prostheses are widely used for the treatment of edentulism, often preferred over the screw-retained ones. However, one of the main features of an implant-supported prosthesis is retrievability, which could be necessary in the case of implant complications. In this study, the retrievability of implant-fixed dental prostheses was investigated considering two of the main factors dental practitioners have to deal with: the abutments geometry and the luting agent. Impulsive forces were applied to dental bridge models to simulate crowns' retrievability in clinical conditions. The number of impulses and the impulsive force delivered during each test were recorded and used as retrievability indexes. One-hundred-and-five tests were conducted on 21 combinations of bridges and luting agents, and a Kruskal-Wallis test was performed on the results. The abutment geometry significantly influenced the number of impulses needed for retrieval ( < 0.05), and a cement-dependent trend was observed as well. On the other hand, the forces measured during tests showed no clear correlation with bridge retrievability. The best retrievability was obtained with long, slightly tapered abutments and a temporary luting agent.

Citing Articles

Evaluation of the dislodgement force of splinted restorations with engaging conical abutments over multiple nonparallel implants (in-vitro study).

Khamis M, Zakaria N, Farrag K, Abolgheit S BMC Oral Health. 2023; 23(1):300.

PMID: 37198623 PMC: 10190041. DOI: 10.1186/s12903-023-02958-4.


The effect of different methods of cleansing temporary cement (with and without eugenol) on the final bond strength of implant-supported zirconia copings after final cementation: An study.

Bajoghli F, Fathi A, Ebadian B, Jowkar M, Sabouhi M Dent Res J (Isfahan). 2023; 20:22.

PMID: 36960017 PMC: 10028587. DOI: 10.4103/1735-3327.369624.


Novel Load Systems for Testing of Biomaterials and Medical Devices.

Bignardi C, Terzini M Materials (Basel). 2023; 16(2).

PMID: 36676202 PMC: 9863571. DOI: 10.3390/ma16020465.


Application of Semipermanent Cements and Conventional Cement with Modified Cementing Technique in Dental Implantology.

Veselinovic V, Marin S, Tatic Z, Trtic N, Dolic O, Adamovic T Acta Stomatol Croat. 2022; 55(4):367-379.

PMID: 35001932 PMC: 8734451. DOI: 10.15644/asc55/4/4.

References
1.
Schiessl C, Schaefer L, Winter C, Fuerst J, Rosentritt M, Zeman F . Factors determining the retentiveness of luting agents used with metal- and ceramic-based implant components. Clin Oral Investig. 2012; 17(4):1179-90. DOI: 10.1007/s00784-012-0798-x. View

2.
Hill E, Lott J . A clinically focused discussion of luting materials. Aust Dent J. 2011; 56 Suppl 1:67-76. DOI: 10.1111/j.1834-7819.2010.01297.x. View

3.
Al Hamad K, Al Rashdan B, Abu-Sitta E . The effects of height and surface roughness of abutments and the type of cement on bond strength of cement-retained implant restorations. Clin Oral Implants Res. 2010; 22(6):638-44. DOI: 10.1111/j.1600-0501.2010.02011.x. View

4.
Heinemann F, Mundt T, Biffar R . Retrospective evaluation of temporary cemented, tooth and implant supported fixed partial dentures. J Craniomaxillofac Surg. 2006; 34 Suppl 2:86-90. DOI: 10.1016/S1010-5182(06)60019-X. View

5.
Mehl C, Harder S, Wolfart M, Kern M, Wolfart S . Retrievability of implant-retained crowns following cementation. Clin Oral Implants Res. 2008; 19(12):1304-11. DOI: 10.1111/j.1600-0501.2008.01587.x. View