Precision and Agreement of Higher Order Aberrations Measured with Ray Tracing and Hartmann-Shack Aberrometers

Overview
Journal BMC Ophthalmol
Publisher Biomed Central
Specialty Ophthalmology
Date 2018 Jan 29
PMID 29374460
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Background: To assess the precision and agreement of measurements of higher order aberrations (HOAs) obtained with a ray tracing aberrometer (iTrace) and a Hartmann-Shack aberrometer (Topcon KR-1 W).

Methods: Prospective evaluation of the diagnostic test. Data from the right eyes of 92 normal subjects obtained using the two devices were included in this study. Two observers performed 3 consecutive scans to determine the intraobserver repeatability and interobserver reproducibility. About one week later, one observer performed an additional 3 consecutive scans to obtain the intersession reproducibility. The within-subject standard deviation (Sw), test-retest repeatability (TRT) and intraclass correlation coefficient (ICC) were used to assess the precision, while Bland-Altman plots were performed to assess the agreement.

Results: For intraobserver repeatability of the ocular, corneal and internal HOAs, Topcon KR-1 W showed a 2.77Sw of 0.079 μm or less and ICCs of 0.761 or more; and iTrace showed a 2.77Sw of 0.105 μm or less and ICCs of 0.805 or more. The ICCs of the internal HOAs of interobserver reproducibility were less than 0.75 except for spherical aberration (SA) (0.862), and interobserver reproducibility of the counterpart showed similar but lower results. For the ocular, corneal and internal HOA measurements, statistically significant differences existed between the Topcon KR-1 W and iTrace (all P < 0.05). No significant differences were observed in the ocular SA and internal coma.

Conclusions: The ray tracing and Hartmann-Shack method aberrometers provided excellent repeatability but less reliable reproducibility in the measurement of HOAs (except for SA). The two aberrometers should not be interchangeable in clinical application because of the significant differences in HOA measurements between them.

Citing Articles

Visual outcomes with implantable Collamer lens versus small incision lenticule extraction in moderate-high myopia: A pilot study.

Sinha R, Daga J, Sahay P, Gupta V, Agarwal T, Sharma N Indian J Ophthalmol. 2024; 73(1):115-121.

PMID: 39257104 PMC: 11831956. DOI: 10.4103/IJO.IJO_908_24.


Efficacy of estimations of Hartmann-Shack sensors in small pupil sizes.

Bartol-Puyal F, Gimenez G, Mendez-Martinez S, Larrosa J, Polo V, Pablo L Saudi J Ophthalmol. 2024; 38(2):168-172.

PMID: 38988783 PMC: 11232754. DOI: 10.4103/sjopt.sjopt_123_23.


Simulation of the Human Myopic Eye Cornea Compensation Based on the Analysis of Aberrometric Data.

Khorin P, Khonina S Vision (Basel). 2023; 7(1).

PMID: 36977301 PMC: 10051312. DOI: 10.3390/vision7010021.


Corneal higher order aberrations by Sirius topography and their relation to different refractive errors.

Salman A, Kailani O, Ghabra M, Omran R, Darwish T, Shaaban R BMC Ophthalmol. 2023; 23(1):104.

PMID: 36927406 PMC: 10018888. DOI: 10.1186/s12886-023-02841-4.


Corneal Higher-Order Aberrations Measurements: Precision of SD-OCT/Placido Topography and Comparison with a Scheimpflug/Placido Topography in Eyes After Small-Incision Lenticule Extraction.

Ning R, Huang X, Jin Y, Lei C, Ma X, Xu S Ophthalmol Ther. 2023; 12(3):1595-1610.

PMID: 36862309 PMC: 10164219. DOI: 10.1007/s40123-023-00693-1.


References
1.
Bland J, Altman D . Measurement error and correlation coefficients. BMJ. 1996; 313(7048):41-2. PMC: 2351452. DOI: 10.1136/bmj.313.7048.41. View

2.
Gabriel C, Klaproth O, Titke C, Baumeister M, Buhren J, Kohnen T . Repeatability of topographic and aberrometric measurements at different accommodative states using a combined topographer and open-view aberrometer. J Cataract Refract Surg. 2015; 41(4):806-11. DOI: 10.1016/j.jcrs.2014.07.037. View

3.
Gobbe M, Reinstein D, Archer T . LASIK-induced aberrations: comparing corneal and whole-eye measurements. Optom Vis Sci. 2015; 92(4):447-55. DOI: 10.1097/OPX.0000000000000557. View

4.
Montes-Mico R, Caliz A, Alio J . Wavefront analysis of higher order aberrations in dry eye patients. J Refract Surg. 2004; 20(3):243-7. DOI: 10.3928/1081-597X-20040501-08. View

5.
Cervino A, Hosking S, Rai G, Naroo S, Gilmartin B . Wavefront analyzers induce instrument myopia. J Refract Surg. 2006; 22(8):795-803. DOI: 10.3928/1081-597X-20061001-10. View