Comparison of Internal and Total Optical Aberrations for 2 Aberrometers: ITrace and OPD Scan

Overview
Specialty Ophthalmology
Date 2008 Dec 20
PMID 19096236
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare and evaluate the total and internal aberrations measured by two aberrometers: the laser ray tracing aberrometer (iTrace, Tracey Technology) and the automatic retinoscope aberrometer (OPD Scan, Nidek).

Methods: A total of 54 healthy eyes were enrolled in the study. Following pupil dilation, aberrations were measured with the iTrace and OPD Scan. We compared the aberrations obtained from measurements obtained at pupillary diameters of 4 mm and 6 mm with the OPD Scan and iTrace. Aberrations of internal optics and total aberrations were compared for the two aberrometers. For each aberrometer and each eye, the averaged Zernike data were used to calculate various root-mean-square (RMS) data. These parameters, together with the refractive parameters, were then analyzed and complimented by paired t-tests.

Results: At a pupil diameter of 4 mm, the number of total aberrations in the entire eye showed significant differences for the mean values of spherical aberrations (Z4,0) obtained with the OPD Scan and iTrace aberrometers (p=0.001). Aberrations of the internal optics showed significant differences in the mean values of total RMS, coma (Z3,-1), and trefoil (Z3,3) between the iTrace and OPD Scan (p<0.001, p=0.01, p<0.001) for the same pupil diameter of 4 mm. At a pupil diameter of 6 mm, the two instruments showed a similar number of total aberrations. Aberrations of the internal optics showed significant differences in the mean values of total RMS, spherical aberration (Z4,0), and coma (Z3,-1) between the two devices (p<0.001, p=0.01, p<0.001).

Conclusions: The iTrace and OPD Scan showed the largest number of differences for aberrations of internal optics rather than total aberrations for both pupil diameters. These results suggest that in healthy eyes, the two aberrometers may vary in some details. The aberrometers showed more agreement at a pupil diameter of 6 mm compared to 4 mm.

Citing Articles

Three-Dimensional Reconstruction and Swept-Source Optical Coherence Tomography for Crystalline Lens Tilt and Decentration Relative to the Corneal Vertex.

Dong J, Wang X, Deng M, Wang X Transl Vis Sci Technol. 2021; 10(9):13.

PMID: 34388238 PMC: 8363763. DOI: 10.1167/tvst.10.9.13.


Changes and Diurnal Variation of Visual Quality after Orthokeratology in Myopic Children.

Guo H, Jin W, Pan A, Wang Q, Qu J, Yu A J Ophthalmol. 2018; 2018:3174826.

PMID: 30410793 PMC: 6205316. DOI: 10.1155/2018/3174826.


From Presbyopia to Cataracts: A Critical Review on Dysfunctional Lens Syndrome.

Fernandez J, Rodriguez-Vallejo M, Martinez J, Tauste A, Pinero D J Ophthalmol. 2018; 2018:4318405.

PMID: 30050689 PMC: 6040261. DOI: 10.1155/2018/4318405.


Precision and agreement of higher order aberrations measured with ray tracing and Hartmann-Shack aberrometers.

Xu Z, Hua Y, Qiu W, Li G, Wu Q BMC Ophthalmol. 2018; 18(1):18.

PMID: 29374460 PMC: 5787271. DOI: 10.1186/s12886-018-0683-8.


Comparative analysis of the visual performance after implantation of the toric implantable collamer lens in stable keratoconus: a 4-year follow-up after sequential procedure (CXL+TICL implantation).

Doroodgar F, Niazi F, Sanginabadi A, Niazi S, Baradaran-Rafii A, Alinia C BMJ Open Ophthalmol. 2018; 2(1):e000090.

PMID: 29354720 PMC: 5721648. DOI: 10.1136/bmjophth-2017-000090.


References
1.
Cheng X, Himebaugh N, Kollbaum P, Thibos L, Bradley A . Test-retest reliability of clinical Shack-Hartmann measurements. Invest Ophthalmol Vis Sci. 2003; 45(1):351-60. DOI: 10.1167/iovs.03-0265. View

2.
Mirshahi A, Buhren J, Gerhardt D, Kohnen T . In vivo and in vitro repeatability of Hartmann-Shack aberrometry. J Cataract Refract Surg. 2004; 29(12):2295-301. DOI: 10.1016/s0886-3350(03)00655-2. View

3.
Rozema J, Van Dyck D, Tassignon M . Clinical comparison of 6 aberrometers. Part 2: statistical comparison in a test group. J Cataract Refract Surg. 2006; 32(1):33-44. DOI: 10.1016/j.jcrs.2004.11.052. View

4.
Liang C, Juo S, Chang C . Comparison of higher-order wavefront aberrations with 3 aberrometers. J Cataract Refract Surg. 2006; 31(11):2153-6. DOI: 10.1016/j.jcrs.2005.04.040. View

5.
Seiler T, Kaemmerer M, Mierdel P, Krinke H . Ocular optical aberrations after photorefractive keratectomy for myopia and myopic astigmatism. Arch Ophthalmol. 2000; 118(1):17-21. DOI: 10.1001/archopht.118.1.17. View