» Articles » PMID: 28671320

Design of Spectral-spatial Phase Prewinding Pulses and Their Use in Small-tip Fast Recovery Steady-state Imaging

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2017 Jul 4
PMID 28671320
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Spectrally selective "prewinding" radiofrequency pulses can counteract B0 inhomogeneity in steady-state sequences, but can only prephase a limited range of off-resonance. We propose spectral-spatial small-tip angle prewinding pulses that increase the off-resonance bandwidth that can be successfully prephased by incorporating spatially tailored excitation patterns.

Theory And Methods: We present a feasibility study to compare spectral and spectral-spatial prewinding pulses. These pulses add a prephasing term to the target magnetization pattern that aims to recover an assigned off-resonance bandwidth at the echo time. For spectral-spatial pulses, the design bandwidth is centered at the off-resonance frequency for each spatial location in a field map. We use these pulses in the small-tip fast recovery steady-state sequence, which is similar to balanced steady-state free precession. We investigate improvement of spectral-spatial pulses over spectral pulses using simulations and small-tip fast recovery scans of a gel phantom and human brain.

Results: In simulation, spectral-spatial pulses yielded lower normalized root mean squared excitation error than spectral pulses. For both experiments, the spectral-spatial pulse images are also qualitatively better (more uniform, less signal loss) than the spectral pulse images.

Conclusion: Spectral-spatial prewinding pulses can prephase over a larger range of off-resonance than their purely spectral counterparts. Magn Reson Med 79:1377-1386, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

References
1.
Mulkern R, Balasubramanian M, Mitsouras D . On the lorentzian versus Gaussian character of time-domain spin-echo signals from the brain as sampled by means of gradient-echoes: Implications for quantitative transverse relaxation studies. Magn Reson Med. 2014; 74(1):51-62. PMC: 4312272. DOI: 10.1002/mrm.25365. View

2.
Bieri O, Scheffler K . SSFP signal with finite RF pulses. Magn Reson Med. 2009; 62(5):1232-41. DOI: 10.1002/mrm.22116. View

3.
Hargreaves B . Rapid gradient-echo imaging. J Magn Reson Imaging. 2012; 36(6):1300-13. PMC: 3502662. DOI: 10.1002/jmri.23742. View

4.
Saekho S, Yip C, Noll D, Boada F, Stenger V . Fast-kz three-dimensional tailored radiofrequency pulse for reduced B1 inhomogeneity. Magn Reson Med. 2006; 55(4):719-24. PMC: 3076290. DOI: 10.1002/mrm.20840. View

5.
Cukur T, Shimakawa A, Yu H, Hargreaves B, Hu B, Nishimura D . Magnetization-prepared IDEAL bSSFP: a flow-independent technique for noncontrast-enhanced peripheral angiography. J Magn Reson Imaging. 2011; 33(4):931-9. PMC: 3155993. DOI: 10.1002/jmri.22479. View