Comparison of Tissue Oximeters on a Liquid Phantom with Adjustable Optical Properties

Overview
Specialty Radiology
Date 2016 Aug 30
PMID 27570691
Citations 51
Authors
Affiliations
Soon will be listed here.
Abstract

The SafeBoosC trial showed that cerebral oximetry combined with a treatment guideline can reduce the the burden of hypoxia in neonates by 50% [Brit. Med. J.350, g7635 (2015)]. However, guidelines based on oximetry by one oximeter are not directly usable by other oximeters. We made a blood-lipid phantom simulating the neonatal head to determine the relation between oxygenation values obtained by different oximeters. We calculated coefficients for easy conversion from one oximeter to the other. We additionally determined the corresponding SafeBoosC intervention thresholds at which we measured an uncertainty of up to 9.2% when varying hemoglobin content from 25μM to 70μM. In conclusion, this paper makes the comparison of absolute values obtained by different oximeters possible.

Citing Articles

Point-of-care optical spectroscopy platform and ratio-metric algorithms for rapid and systematic functional characterization of biological models .

Hasan M, Yan J, Zhu C J Biomed Opt. 2025; 29(12):125002.

PMID: 39741576 PMC: 11687633. DOI: 10.1117/1.JBO.29.12.125002.


Light-guided dynamic phantom to mimic microvasculature for biomedical applications: an exploration for pulse oximeter.

Ma H, Angelone D, Guadagno C, Andersson-Engels S, Konugolu Venkata Sekar S J Biomed Opt. 2024; 29(Suppl 3):S33312.

PMID: 39697671 PMC: 11654752. DOI: 10.1117/1.JBO.29.S3.S33312.


Neonatal somatic oxygenation and perfusion assessment using near-infrared spectroscopy : Part of the series on near-infrared spectroscopy by the European Society of Paediatric Research Special Interest Group "Near-Infrared Spectroscopy".

Kooi E, Mintzer J, Rhee C, Ergenekon E, Schwarz C, Pichler G Pediatr Res. 2024; 96(5):1180-1194.

PMID: 38730022 DOI: 10.1038/s41390-024-03226-z.


The influence of carbon dioxide on cerebral metabolism and oxygen consumption: combining multimodal monitoring with dynamic systems modelling.

Highton D, Caldwell M, Tachtsidis I, Elwell C, Smith M, Cooper C Biol Open. 2024; 13(1).

PMID: 38180242 PMC: 10810564. DOI: 10.1242/bio.060087.


Comparison of the SenSmart™ and the INVOS™ neonatal cerebral near-infrared spectrometry devices.

Wirayannawat W, Amawat J, Yamsiri N, Paes B, Kitsommart R Front Pediatr. 2023; 11:1243977.

PMID: 37691777 PMC: 10485771. DOI: 10.3389/fped.2023.1243977.


References
1.
De Grand A, Lomnes S, Lee D, Pietrzykowski M, Ohnishi S, Morgan T . Tissue-like phantoms for near-infrared fluorescence imaging system assessment and the training of surgeons. J Biomed Opt. 2006; 11(1):014007. PMC: 2486408. DOI: 10.1117/1.2170579. View

2.
Nosoh Y . ABSORPTION SPECTRUM OF ACTIVELY RESPIRING YEAST CELLS. Arch Biochem Biophys. 1964; 105:439-45. DOI: 10.1016/0003-9861(64)90028-1. View

3.
Nasseri N, Kleiser S, Reidt S, Wolf M . Local Measurement of Flap Oxygen Saturation: An Application of Visible Light Spectroscopy. Adv Exp Med Biol. 2016; 876:391-397. DOI: 10.1007/978-1-4939-3023-4_49. View

4.
Yoshida T, Shevkoplyas S . Anaerobic storage of red blood cells. Blood Transfus. 2010; 8(4):220-36. PMC: 2957487. DOI: 10.2450/2010.0022-10. View

5.
Kleiser S, Hyttel-Sorensen S, Greisen G, Wolf M . Comparison of Near-Infrared Oximeters in a Liquid Optical Phantom with Varying Intralipid and Blood Content. Adv Exp Med Biol. 2016; 876:413-418. DOI: 10.1007/978-1-4939-3023-4_52. View