» Articles » PMID: 12791110

Measurement of Cerebral Oxygenation State in Anaesthetized Children Using the INVOS 5100 Cerebral Oximeter

Overview
Date 2003 Jun 7
PMID 12791110
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Near-infrared spectroscopy is a developing technique for monitoring cerebral oxygenation during anaesthesia. The aim of this study was to evaluate absolute values of cerebral oxygenation during stable anaesthesia conditions in otherwise healthy children using the recently introduced INVOS 5100 cerebral oximeter with a paediatric and adult sensor and to compare them with values obtained from the NIRO 300 oximeter.

Methods: Thirty paediatric surgical patients (aged 0.23-15.97 years) were studied during general anaesthesia with tracheal intubation and controlled ventilation. Comparative measurements of cerebral oxygenation were performed on the forehead with two probes within 10 min under stable cardiorespiratory and anaesthesia conditions. Cerebral oxygenation values (rSO2) obtained from the paediatric and adult INVOS 5100 sensors were compared with the tissue oxygenation index (TOI) obtained from the NIRO 300 cerebral oximeter using 4- and 5-cm emitter-detector separation.

Results: Cerebral rSO2 values and the TOI values both showed a large range of cerebral oxygenation in the children studied (rSO2: 59-95%, TOI: 48-85%). Cerebral rSO2 values measured by the INVOS 5100, particularly with the paediatric sensor, were significantly higher than the TOI values obtained from the NIRO 300 (P < 0.0001). Agreement between the INVOS and NIRO oximeter was poor.

Conclusion: The large range and the poor agreement of cerebral oxygenation values between the two oximeters makes it difficult to define a normal value. Cerebral oxygenation readings by these monitors, based on one single point measurement during anaesthesia, should be viewed with caution. Actually, there may be little indication for routine use of such monitoring during general anaesthesia.

Citing Articles

Reference values for cerebral and peripheral tissue oximetry in children: A clinical TD-NIRS study.

Calcaterra V, Lacerenza M, Rossi V, Zanelli S, Contini D, Amendola C Acta Paediatr. 2024; 114(3):515-525.

PMID: 39425553 PMC: 11828717. DOI: 10.1111/apa.17459.


Cerebral oxygenation saturation in childhood: difference by age and comparison of two cerebral oximetry algorithms.

Kubo Y, Itosu Y, Kubo T, Saito H, Okada K, Ito Y J Clin Monit Comput. 2024; 38(3):639-648.

PMID: 38310594 DOI: 10.1007/s10877-023-01124-z.


Could Near Infrared Spectroscopy (NIRS) be the new weapon in our fight against Necrotising Enterocolitis?.

Howarth C, Banerjee J, Leung T, Aladangady N Front Pediatr. 2022; 10:1024566.

PMID: 36425397 PMC: 9679512. DOI: 10.3389/fped.2022.1024566.


Three Physiological Components That Influence Regional Cerebral Tissue Oxygen Saturation.

Suppan E, Pichler G, Binder-Heschl C, Schwaberger B, Urlesberger B Front Pediatr. 2022; 10:913223.

PMID: 35769216 PMC: 9234387. DOI: 10.3389/fped.2022.913223.


Absolute quantification of cerebral tissue oxygen saturation with multidistance broadband NIRS in newborn brain.

Kovacsova Z, Bale G, Mitra S, Lange F, Tachtsidis I Biomed Opt Express. 2021; 12(2):907-925.

PMID: 33680549 PMC: 7901317. DOI: 10.1364/BOE.412088.