Multisource Inverse-geometry CT. Part I. System Concept and Development

Overview
Journal Med Phys
Specialty Biophysics
Date 2016 Aug 5
PMID 27487877
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: This paper presents an overview of multisource inverse-geometry computed tomography (IGCT) as well as the development of a gantry-based research prototype system. The development of the distributed x-ray source is covered in a companion paper [V. B. Neculaes et al., "Multisource inverse-geometry CT. Part II. X-ray source design and prototype," Med. Phys. 43, 4617-4627 (2016)]. While progress updates of this development have been presented at conferences and in journal papers, this paper is the first comprehensive overview of the multisource inverse-geometry CT concept and prototype. The authors also provide a review of all previous IGCT related publications.

Methods: The authors designed and implemented a gantry-based 32-source IGCT scanner with 22 cm field-of-view, 16 cm z-coverage, 1 s rotation time, 1.09 × 1.024 mm detector cell size, as low as 0.4 × 0.8 mm focal spot size and 80-140 kVp x-ray source voltage. The system is built using commercially available CT components and a custom made distributed x-ray source. The authors developed dedicated controls, calibrations, and reconstruction algorithms and evaluated the system performance using phantoms and small animals.

Results: The authors performed IGCT system experiments and demonstrated tube current up to 125 mA with up to 32 focal spots. The authors measured a spatial resolution of 13 lp/cm at 5% cutoff. The scatter-to-primary ratio is estimated 62% for a 32 cm water phantom at 140 kVp. The authors scanned several phantoms and small animals. The initial images have relatively high noise due to the low x-ray flux levels but minimal artifacts.

Conclusions: IGCT has unique benefits in terms of dose-efficiency and cone-beam artifacts, but comes with challenges in terms of scattered radiation and x-ray flux limits. To the authors' knowledge, their prototype is the first gantry-based IGCT scanner. The authors summarized the design and implementation of the scanner and the authors presented results with phantoms and small animals.

Citing Articles

A simulated comparison of lung tumor target verification using stereoscopic tomosynthesis or radiography.

Hsieh S, Ng L, Cao M, Lee P Med Phys. 2022; 49(5):3041-3052.

PMID: 35319790 PMC: 10471465. DOI: 10.1002/mp.15634.


Tetrahedron beam imaging with a multi-pixel thermionic emission X-ray source and a photon-counting detector: A benchtop experimental study.

Chen Q, Zhou S, Tan Y, Hao Y, Zhang Z, Mazur T Med Phys. 2021; 48(11):7250-7260.

PMID: 34482562 PMC: 8595623. DOI: 10.1002/mp.15211.


Multisource inverse-geometry CT. Part II. X-ray source design and prototype.

Neculaes V, Caiafa A, Cao Y, De Man B, Edic P, Frutschy K Med Phys. 2016; 43(8):4617.

PMID: 27487878 PMC: 4958104. DOI: 10.1118/1.4954847.

References
1.
Schmidt T, Fahrig R, Pelc N, Solomon E . An inverse-geometry volumetric CT system with a large-area scanned source: a feasibility study. Med Phys. 2004; 31(9):2623-7. DOI: 10.1118/1.1786171. View

2.
Tang X, Hsieh J, Hagiwara A, Nilsen R, Thibault J, Drapkin E . A three-dimensional weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT under a circular source trajectory. Phys Med Biol. 2005; 50(16):3889-905. DOI: 10.1088/0031-9155/50/16/016. View

3.
Flohr T, McCollough C, Bruder H, Petersilka M, Gruber K, Suss C . First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol. 2005; 16(2):256-68. DOI: 10.1007/s00330-005-2919-2. View

4.
Schmidt T, Fahrig R, Pelc N . A three-dimensional reconstruction algorithm for an inverse-geometry volumetric CT system. Med Phys. 2005; 32(11):3234-45. DOI: 10.1118/1.2064827. View

5.
Schmidt T, Star-Lack J, Bennett N, Mazin S, Solomon E, Fahrig R . A prototype table-top inverse-geometry volumetric CT system. Med Phys. 2006; 33(6):1867-78. DOI: 10.1118/1.2192887. View