A Multi-source Inverse-geometry CT System: Initial Results with an 8 Spot X-ray Source Array

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2014 Feb 22
PMID 24556567
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

We present initial experimental results of a rotating-gantry multi-source inverse-geometry CT (MS-IGCT) system. The MS-IGCT system was built with a single module of 2 × 4 x-ray sources and a 2D detector array. It produced a 75 mm in-plane field-of-view (FOV) with 160 mm axial coverage in a single gantry rotation. To evaluate system performance, a 2.5 inch diameter uniform PMMA cylinder phantom, a 200 µm diameter tungsten wire, and a euthanized rat were scanned. Each scan acquired 125 views per source and the gantry rotation time was 1 s per revolution. Geometric calibration was performed using a bead phantom. The scanning parameters were 80 kVp, 125 mA, and 5.4 µs pulse per source location per view. A data normalization technique was applied to the acquired projection data, and beam hardening and spectral nonlinearities of each detector channel were corrected. For image reconstruction, the projection data of each source row were rebinned into a full cone beam data set, and the FDK algorithm was used. The reconstructed volumes from upper and lower source rows shared an overlap volume which was combined in image space. The images of the uniform PMMA cylinder phantom showed good uniformity and no apparent artifacts. The measured in-plane MTF showed 13 lp cm(-1) at 10% cutoff, in good agreement with expectations. The rat data were also reconstructed reliably. The initial experimental results from this rotating-gantry MS-IGCT system demonstrated its ability to image a complex anatomical object without any significant image artifacts and to achieve high image resolution and large axial coverage in a single gantry rotation.

Citing Articles

Tetrahedron beam imaging with a multi-pixel thermionic emission X-ray source and a photon-counting detector: A benchtop experimental study.

Chen Q, Zhou S, Tan Y, Hao Y, Zhang Z, Mazur T Med Phys. 2021; 48(11):7250-7260.

PMID: 34482562 PMC: 8595623. DOI: 10.1002/mp.15211.


Single-view geometric calibration for C-arm inverse geometry CT.

Slagowski J, Dunkerley D, Hatt C, Speidel M J Med Imaging (Bellingham). 2017; 4(1):013506.

PMID: 28560241 PMC: 5358550. DOI: 10.1117/1.JMI.4.1.013506.


Multisource inverse-geometry CT. Part I. System concept and development.

De Man B, Uribe J, Baek J, Harrison D, Yin Z, Longtin R Med Phys. 2016; 43(8):4607.

PMID: 27487877 PMC: 4958105. DOI: 10.1118/1.4954846.


Raw data normalization for a multi source inverse geometry CT system.

Baek J, De Man B, Harrison D, Pelc N Opt Express. 2015; 23(6):7514-26.

PMID: 25837090 PMC: 4408891. DOI: 10.1364/OE.23.007514.

References
1.
Baek J, Pelc N . The noise power spectrum in CT with direct fan beam reconstruction. Med Phys. 2010; 37(5):2074-81. PMC: 2871306. DOI: 10.1118/1.3378673. View

2.
Kohl G . The evolution and state-of-the-art principles of multislice computed tomography. Proc Am Thorac Soc. 2005; 2(6):470-6, 499-500. DOI: 10.1513/pats.200508-086DS. View

3.
Nickoloff E, Riley R . A simplified approach for modulation transfer function determinations in computed tomography. Med Phys. 1985; 12(4):437-42. DOI: 10.1118/1.595706. View

4.
Schmidt T, Fahrig R, Pelc N, Solomon E . An inverse-geometry volumetric CT system with a large-area scanned source: a feasibility study. Med Phys. 2004; 31(9):2623-7. DOI: 10.1118/1.1786171. View

5.
Fishman E, Horton K . Imaging pancreatic cancer: the role of multidetector CT with three-dimensional CT angiography. Pancreatology. 2002; 1(6):610-24. DOI: 10.1159/000055871. View