Clinical Evaluation of an Unsintered Hydroxyapatite/Poly-L-Lactide Osteoconductive Composite Device for the Internal Fixation of Maxillofacial Fractures

Overview
Date 2016 Jul 19
PMID 27428913
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: OSTEOTRANS MX (Takiron Co, Ltd, Osaka, Japan) is a resorbable osteosynthetic material composed of an unsintered hydroxyapatite/poly-L-lactide composite, and its osteoconductive capacity has been documented. The authors here report their clinical experience using OSTEOTRANS MX.

Methods: The authors treated 35 patients (19 men, 16 women; age, 14-88 years; mean ± standard deviation, 38.4 ± 19.9 years) with maxillofacial fractures. The authors used standard surgery to stabilize fractures in all patients, fitting resorbable plates (thickness, 1.0 or 1.4 mm) and screws (diameter, 2 mm) according to Arbeitsgemeinschaft für Osteosynthesefragen/Association (AO) for the Study of Internal Fixation guidelines.

Results: All patients eventually achieved satisfactory healing with favorable restoration of form and function without foreign body reaction. Complications occurred in 3 patients-plate exposure in 2 and discomfort in 1. However, fracture sites healed in all patients. Scanning electron microscopy revealed that the devices bonded directly to the bone without interposition of nonmineralized tissue.

Conclusion: OSTEOTRANS MX is a useful material with few complications. Its osteoconductive bioactivity is advantageous for the early functional improvement of maxillofacial fractures.

Citing Articles

Biomechanical Evaluation of Hydroxyapatite/poly-l-lactide Fixation in Mandibular Body Reconstruction with Fibula Free Flap: A Finite Element Analysis Incorporating Material Properties and Masticatory Function Evaluation.

Lee S, Jung T, Kim W, Kim B, Lee J Bioengineering (Basel). 2024; 11(10).

PMID: 39451385 PMC: 11505077. DOI: 10.3390/bioengineering11101009.


Engineering of Bioresorbable Polymers for Tissue Engineering and Drug Delivery Applications.

Dobrzynska-Mizera M, Dodda J, Liu X, Knitter M, Oosterbeek R, Salinas P Adv Healthc Mater. 2024; 13(30):e2401674.

PMID: 39233521 PMC: 11616265. DOI: 10.1002/adhm.202401674.


Cephalometric evaluation of skeletal stability and pharyngeal airway changes after mandibular setback surgery: Bioabsorbable versus titanium plate and screw fixation.

Thet P, Kaboosaya B Imaging Sci Dent. 2024; 54(2):181-190.

PMID: 38948194 PMC: 11211026. DOI: 10.5624/isd.20230269.


Comparative In Vitro Dissolution Assessment of Calcined and Uncalcined Hydroxyapatite Using Differences in Bioresorbability and Biomineralization.

Jang W, Pyun J, Chang J Int J Mol Sci. 2024; 25(1).

PMID: 38203791 PMC: 10779548. DOI: 10.3390/ijms25010621.


Evaluation of Hard and Soft Tissue Responses to Four Different Generation Bioresorbable Materials-Poly-l-Lactic Acid (PLLA), Poly-l-Lactic Acid/Polyglycolic Acid (PLLA/PGA), Uncalcined/Unsintered Hydroxyapatite/Poly-l-Lactic Acid (u-HA/PLLA) and....

Ayasaka K, Ramanathan M, Huy N, Shijirbold A, Okui T, Tatsumi H Materials (Basel). 2023; 16(23).

PMID: 38068124 PMC: 10707338. DOI: 10.3390/ma16237379.


References
1.
Landes C, Ballon A, Tran A, Ghanaati S, Sader R . Segmental stability in orthognathic surgery: hydroxyapatite/Poly-l-lactide osteoconductive composite versus titanium miniplate osteosyntheses. J Craniomaxillofac Surg. 2014; 42(6):930-42. DOI: 10.1016/j.jcms.2014.01.013. View

2.
Tormala P, Vasenius J, Vainionpaa S, Laiho J, Pohjonen T, Rokkanen P . Ultra-high-strength absorbable self-reinforced polyglycolide (SR-PGA) composite rods for internal fixation of bone fractures: in vitro and in vivo study. J Biomed Mater Res. 1991; 25(1):1-22. DOI: 10.1002/jbm.820250102. View

3.
Bostman O, Hirvensalo E, Makinen J, Rokkanen P . Foreign-body reactions to fracture fixation implants of biodegradable synthetic polymers. J Bone Joint Surg Br. 1990; 72(4):592-6. DOI: 10.1302/0301-620X.72B4.2199452. View

4.
Kubota Y, Kuroki T, Akita S, Koizumi T, Hasegawa M, Rikihisa N . Association between plate location and plate removal following facial fracture repair. J Plast Reconstr Aesthet Surg. 2011; 65(3):372-8. DOI: 10.1016/j.bjps.2011.09.040. View

5.
Esen A, Ataoglu H, Gemi L . Comparison of stability of titanium and absorbable plate and screw fixation for mandibular angle fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008; 106(6):806-11. DOI: 10.1016/j.tripleo.2008.05.034. View