Changes in the U.S. Primary Standards for the Air Kerma From Gamma-Ray Beams

Overview
Date 2016 Jul 15
PMID 27413615
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Monte Carlo photon-electron transport calculations have been done to derive new wall corrections for the six NBS-NIST standard graphite-wall, air-ionization cavity chambers that serve as the U.S. national primary standard for air kerma (and exposure) for gamma rays from (60)Co, (137)Cs, and (192)Ir sources. The data developed for and from these calculations have also been used to refine a number of other factors affecting the standards. The largest changes are due to the new wall corrections, and the total changes are +0.87 % to +1.11 % (depending on the chamber) for (60)Co beams, +0.64 % to +1.07 % (depending on the chamber) for (137)Cs beams, and -0.06 % for the single chamber used in the measurement of the standardized (192)Ir source. The primary standards for air kerma will be adjusted in the near future to reflect the changes in factors described in this work.

Citing Articles

New National Air-Kerma-Strength Standards for (125)I and (103)Pd Brachytherapy Seeds.

Seltzer S, Lamperti P, Loevinger R, Mitch M, Weaver J, Coursey B J Res Natl Inst Stand Technol. 2016; 108(5):337-58.

PMID: 27413614 PMC: 4847582. DOI: 10.6028/jres.108.030.


Evaluation of wall correction factor of INER's air-kerma primary standard chamber and dose variation by source displacement for HDR ¹⁹²Ir brachytherapy.

Lee J, Wang J, Huang T, Su S, Chang B, Su C Biomed Res Int. 2013; 2013:436979.

PMID: 24222907 PMC: 3816024. DOI: 10.1155/2013/436979.

References
1.
Loftus T, Weaver J . Standardization of Co and Cs Gamma-Ray Beams in Terms of Exposure. J Res Natl Bur Stand A Phys Chem. 2020; 78A(4):465-476. PMC: 6742815. DOI: 10.6028/jres.078A.030. View

2.
. Report 85: Fundamental quantities and units for ionizing radiation. J ICRU. 2013; 11(1):1-31. DOI: 10.1093/jicru/ndr011. View

3.
Borg J, Kawrakow I, Rogers D, Seuntjens J . Monte Carlo study of correction factors for Spencer-Attix cavity theory at photon energies at or above 100 keV. Med Phys. 2000; 27(8):1804-13. DOI: 10.1118/1.1287054. View

4.
Nahum A . Water/air mass stopping power ratios for megavoltage photon and electron beams. Phys Med Biol. 1978; 23(1):24-38. DOI: 10.1088/0031-9155/23/1/002. View

5.
Seltzer S . Calculation of photon mass energy-transfer and mass energy-absorption coefficients. Radiat Res. 1993; 136(2):147-70. View