» Articles » PMID: 27413614

New National Air-Kerma-Strength Standards for (125)I and (103)Pd Brachytherapy Seeds

Overview
Date 2016 Jul 15
PMID 27413614
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

The new U.S. measurement standard for the air-kerma strength from low-energy photon-emitting brachytherapy seed sources is formally described in detail. This instrument-based standard was implemented on 1 January 1999, with its salient features and the implications of differences with the previous standard given only through a series of informal communications. The Wide-Angle Free-Air Chamber (WAFAC) is specially designed to realize air kerma from a single-seed source emitting photons with energies up to about 40 keV, and is now used to measure the wide variety of seeds used in prostate-cancer therapy that has appeared in the last few years. For the two (125)I seed models that have been subject to both the old and new standards, the new standard reduces the air-kerma strength by 10.3 %. This change is mainly due to the removal of the influence on the measurement of the Ti K x rays produced in the source encapsulation, a component with no clinical significance.

Citing Articles

Dosimetric evaluation of incorporating the revised V4.0 calibration protocol for breast intraoperative radiotherapy with the INTRABEAM system.

Shaikh M, Burmeister J, Scott R, Kumaraswamy L, Nalichowski A, Joiner M J Appl Clin Med Phys. 2020; 21(2):50-59.

PMID: 32039545 PMC: 7020998. DOI: 10.1002/acm2.12807.


Dosimetric characterization of a new directional low-dose rate brachytherapy source.

Aima M, DeWerd L, Mitch M, Hammer C, Culberson W Med Phys. 2018; .

PMID: 29797517 PMC: 6548702. DOI: 10.1002/mp.12994.


Air-kerma strength determination of a new directional (103)Pd source.

Aima M, Reed J, DeWerd L, Culberson W Med Phys. 2015; 42(12):7144-52.

PMID: 26632069 PMC: 5148138. DOI: 10.1118/1.4935409.


New National Air-Kerma Standard for Low-Energy Electronic Brachytherapy Sources.

Seltzer S, OBrien M, Mitch M J Res Natl Inst Stand Technol. 2015; 119:554-74.

PMID: 26601044 PMC: 4487278. DOI: 10.6028/jres.119.022.


Air kerma and absorbed dose standards for reference dosimetry in brachytherapy.

Sander T Br J Radiol. 2014; 87(1041):20140176.

PMID: 24814696 PMC: 4453150. DOI: 10.1259/bjr.20140176.


References
1.
Loftus T, Weaver J . Standardization of Co and Cs Gamma-Ray Beams in Terms of Exposure. J Res Natl Bur Stand A Phys Chem. 2020; 78A(4):465-476. PMC: 6742815. DOI: 10.6028/jres.078A.030. View

2.
Loftus T . Exposure Standardization of Iodine-125 Seeds Used for Brachytherapy. J Res Natl Bur Stand (1977). 2021; 89(4):295-303. PMC: 6768197. DOI: 10.6028/jres.089.017. View

3.
Ritz V . Design of free-air ionization chambers for the soft x-ray region (20-100 kv). Radiology. 1959; 73:911-22. DOI: 10.1148/73.6.911. View

4.
Loftus T . Standardization of Cesium-137 Gamma-Ray Sources in Terms of Exposure Units (Roentgens). J Res Natl Bur Stand A Phys Chem. 2020; 74A(1):1-6. PMC: 6753037. DOI: 10.6028/jres.074A.001. View

5.
Seltzer S, Bergstrom Jr P . Changes in the U.S. Primary Standards for the Air Kerma From Gamma-Ray Beams. J Res Natl Inst Stand Technol. 2016; 108(5):359-81. PMC: 4847583. DOI: 10.6028/jres.108.031. View