Monte Carlo Simulation of a Typical 60Co Therapy Source

Overview
Journal Med Phys
Specialty Biophysics
Date 1999 Dec 10
PMID 10587239
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

The BEAM Monte Carlo code is used to simulate the 60Co beam from an Eldorado 6 radiotherapy unit and to calculate the relative air-kerma output factors as a function of field size. The unit is realistically modeled, including source capsule, housing and collimator assembly. The calculated relative air-kerma output factors at SSD=80.5 cm agree to within 0.1% with measured values. It is shown that the variation of the output factor is almost entirely due to scattered photons from the fixed and adjustable collimators and there is no effect of shadowing primary photons. The influence of the geometry of the collimation system on the photon spectra on-axis is shown to be small but finite. The calculated buildup region of a depth-dose curve in a water phantom irradiated by a narrow and a broad 60Co beam is shown to agree with experimental data at the 2% to 3% level. Unlike previous calculations, the results accurately predict the effects of electron contamination from the surface to dose maximum. The variation of electron contamination with field size is also presented, as are spectra as a function of field size.

Citing Articles

Beam quality correction factors for dose measurements around Ir brachytherapy sources.

Thrapsanioti Z, Peppa V, Hourdakis C, Karaiskos P, Lekatou A, Pantelis E J Appl Clin Med Phys. 2024; 26(1):e14575.

PMID: 39601314 PMC: 11713369. DOI: 10.1002/acm2.14575.


Technical note: Consistency of IAEA's TRS-483 and AAPM's extended TG-51 protocols for clinical reference dosimetry of the CyberKnife M6 machine.

Duchaine J, Markel D, Ley J, Beliveau-Nadeau D, Zerouali K, Doucet R J Appl Clin Med Phys. 2023; 24(5):e13976.

PMID: 36995902 PMC: 10161123. DOI: 10.1002/acm2.13976.


Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities.

Pantelis E, Logothetis A, Zoros E, Pappas E, Papagiannis P, Paddick I J Appl Clin Med Phys. 2023; 24(5):e13903.

PMID: 36655619 PMC: 10161100. DOI: 10.1002/acm2.13903.


Determination of the dose rate around a HDR Ir brachytherapy source with the microDiamond and the microSilicon detector.

Rossi G, Failing T, Gainey M, Kollefrath M, Hensley F, Zink K Z Med Phys. 2022; 33(4):463-478.

PMID: 36038432 PMC: 10751698. DOI: 10.1016/j.zemedi.2022.07.004.


Enhancement of the EGSnrc code egs_chamber for fast fluence calculations of charged particles.

Failing T, Hartmann G, Hensley F, Keil B, Zink K Z Med Phys. 2022; 32(4):417-427.

PMID: 35643800 PMC: 9948836. DOI: 10.1016/j.zemedi.2022.04.003.