TG-43U1 Parameterization of Elongated RadioCoil 103Pd Brachytherapy Sources

Overview
Date 2007 Aug 23
PMID 17712301
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Recently, to eliminate problems associated with seed type sources, RadioMed Corporation (Tyngsboro, MA) introduced RadioCoil 103Pd sources for interstitial prostate implants. The RadioCoil sources are available in integral lengths ranging from 1.0 cm to 6.0 cm. In this project, dosimetric characteristics of these sources were determined following the TG-43U1 recommendations, with consideration of our recent publication on the evaluation of two-dimensional anisotropy function for elongated brachytherapy sources. Dosimetric parameters of these sources were determined experimentally in Solid Water (Gammex RMI, Middleton, WI) and theoretically using Monte Carlo simulation in Solid Water and liquid water. Per the TG-43U1 protocol, the consensus of these results would be used for their clinical applications.

Citing Articles

Evaluation of hypothetical (153)Gd source for use in brachytherapy.

Ghorbani M, Behmadi M Rep Pract Oncol Radiother. 2016; 21(1):17-24.

PMID: 26900353 PMC: 4716406. DOI: 10.1016/j.rpor.2015.05.005.


Thermoluminescent and Monte Carlo dosimetry of IR06-103Pd brachytherapy source.

Saidi P, Sadeghi M, Hosseini S, Tenreiro C J Appl Clin Med Phys. 2011; 12(4):3581.

PMID: 22089014 PMC: 5718737. DOI: 10.1120/jacmp.v12i4.3581.


Cylindrical coordinate based TG-43U1 parameters for dose calculation around elongated brachytherapy sources.

Awan S, Dini S, Hussain M, Soleimani-Meigooni D, S Meigooni A J Appl Clin Med Phys. 2008; 9(2):123-142.

PMID: 18714281 PMC: 5721703. DOI: 10.1120/jacmp.v9i2.2760.

References
1.
Seltzer S, Lamperti P, Loevinger R, Mitch M, Weaver J, Coursey B . New National Air-Kerma-Strength Standards for (125)I and (103)Pd Brachytherapy Seeds. J Res Natl Inst Stand Technol. 2016; 108(5):337-58. PMC: 4847582. DOI: 10.6028/jres.108.030. View

2.
Yu Y, Waterman F, SUNTHARALINGAM N, Schulsinger A . Limitations of the minimum peripheral dose as a parameter for dose specification in permanent 125I prostate implants. Int J Radiat Oncol Biol Phys. 1996; 34(3):717-25. DOI: 10.1016/0360-3016(95)02153-1. View

3.
Melhus C, Rivard M . Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources. Med Phys. 2006; 33(6):1729-37. DOI: 10.1118/1.2199987. View

4.
S Meigooni A, Hayes J, Zhang H, Sowards K . Experimental and theoretical determination of dosimetric characteristics of IsoAid ADVANTAGE 125I brachytherapy source. Med Phys. 2002; 29(9):2152-8. DOI: 10.1118/1.1500395. View

5.
Duggan D . Improved radial dose function estimation using current version MCNP Monte-Carlo simulation: Model 6711 and ISC3500 125I brachytherapy sources. Appl Radiat Isot. 2004; 61(6):1443-50. DOI: 10.1016/j.apradiso.2004.05.070. View