Scientific Basis for Standardization of Fetal Head Measurements by Ultrasound: a Reproducibility Study

Overview
Date 2016 May 10
PMID 27158767
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare the standard methods for ultrasound measurement of fetal head circumference (HC) and biparietal diameter (BPD) (outer-to-outer (BPDoo) vs outer-to-inner (BPDoi) caliper placement), and compare acquisition of these measurements in transthalamic (TT) vs transventricular (TV) planes.

Methods: This study utilized ultrasound images acquired from women participating in the Oxford arm of the INTERGROWTH-21(st) Project. In the first phase of the study, BPDoo and BPDoi were measured on stored images. In the second phase, real-time measurements of BPD, occipitofrontal diameter (OFD) and HC in TT and TV planes were obtained by pairs of sonographers. Reproducibility of measurements made by the same (intraobserver) and by different (interobserver) sonographers, as well as the reproducibility of caliper placement and measurements obtained in different planes, was assessed using Bland-Altman plots.

Results: In Phase I, we analyzed ultrasound images of 108 singleton fetuses. The mean intraobserver and interobserver differences were < 2% (1.34 mm) and the 95% limits of agreement were < 5% (3 mm) for both BPDoo and BPDoi. Neither method for measuring BPD showed consistently better reproducibility. In Phase II, we analyzed ultrasound images of 100 different singleton fetuses. The mean intraobserver and interobserver differences were < 1% (2.26 mm) and the 95% limits of agreement were < 8% (14.45 mm) for all fetal head measurements obtained in TV and TT planes. Neither plane for measuring fetal head showed consistently better reproducibility. Measurement of HC using the ellipse facility was as reproducible as HC calculated from BPD and OFD. OFD by itself was the least reproducible of all fetal head measurements.

Conclusions: Measurements of BPDoi and BPDoo are equally reproducible; however, we believe BPDoo should be used in clinical practice as it allows fetal HC to be measured and compared with neonatal HC. For all head measurements, TV and TT planes provide equally reproducible values at any gestational age, and HC values are similar in both planes. Fetal head measurement in the TT plane is preferable as international standards in this plane are available; however, measurements in the TV plane can be plotted on the same standards. Copyright © 2016 ISUOG. Published by John Wiley & Sons Ltd.

Citing Articles

Transvaginal cervical cerclage: double monofilament modified Wurm vs single braided McDonald technique.

Donadono V, Koutikwar P, Banerjee A, Ivan M, Colley C, Sciacca M Ultrasound Obstet Gynecol. 2025; 65(3):344-352.

PMID: 39998160 PMC: 11872348. DOI: 10.1002/uog.29184.


Modality-level obstacles and initiatives to improve representation in fetal, infant, and toddler neuroimaging research samples.

Margolis E, Nelson P, Fiske A, Champaud J, Olson H, Gomez M Dev Cogn Neurosci. 2025; 72:101505.

PMID: 39954600 PMC: 11875194. DOI: 10.1016/j.dcn.2024.101505.


Reproducibility of fetal ultrasound doppler parameters used for growth assessment.

Zidan Sweid R, Donadono V, Casagrandi D, Sarno L, Attilakos G, Pandya P Arch Gynecol Obstet. 2025; .

PMID: 39821448 DOI: 10.1007/s00404-024-07883-7.


Three-Dimensional Volume Ultrasound Assessment of Cesarean Scar Niche and Cervix in Pregnant Women: A Reproducibility Study.

Ivan M, Mahdy H, Banerjee A, Tetteh A, Greenwold N, Casagrandi D J Ultrasound Med. 2024; 44(3):509-519.

PMID: 39508476 PMC: 11796327. DOI: 10.1002/jum.16613.


Etiologically Based Functional Taxonomy of the Preterm Birth Syndrome.

Villar J, Cavoretto P, Barros F, Romero R, Papageorghiou A, Kennedy S Clin Perinatol. 2024; 51(2):475-495.

PMID: 38705653 PMC: 11632914. DOI: 10.1016/j.clp.2024.02.014.


References
1.
Gull I, Fait G, Har-Toov J, Kupferminc M, Lessing J, Jaffa A . Prediction of fetal weight by ultrasound: the contribution of additional examiners. Ultrasound Obstet Gynecol. 2002; 20(1):57-60. DOI: 10.1046/j.1469-0705.2002.00742.x. View

2.
Chan L, Fung T, Leung T, Sahota D, Lau T . Volumetric (3D) imaging reduces inter- and intraobserver variation of fetal biometry measurements. Ultrasound Obstet Gynecol. 2009; 33(4):447-52. DOI: 10.1002/uog.6321. View

3.
Hadlock F, Deter R, Harrist R, Park S . Fetal head circumference: relation to menstrual age. AJR Am J Roentgenol. 1982; 138(4):649-53. DOI: 10.2214/ajr.138.4.649. View

4.
Sarris I, Ioannou C, Chamberlain P, Ohuma E, Roseman F, Hoch L . Intra- and interobserver variability in fetal ultrasound measurements. Ultrasound Obstet Gynecol. 2012; 39(3):266-73. DOI: 10.1002/uog.10082. View

5.
Ioannou C, Javaid M, Mahon P, Yaqub M, Harvey N, Godfrey K . The effect of maternal vitamin D concentration on fetal bone. J Clin Endocrinol Metab. 2012; 97(11):E2070-7. PMC: 3485609. DOI: 10.1210/jc.2012-2538. View