» Articles » PMID: 27026243

Cranial Fixation Plates in Cerebral Magnetic Resonance Imaging: a 3 and 7 Tesla in Vivo Image Quality Study

Overview
Journal MAGMA
Publisher Springer
Date 2016 Mar 31
PMID 27026243
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: This study assesses and quantifies impairment of postoperative magnetic resonance imaging (MRI) at 7 Tesla (T) after implantation of titanium cranial fixation plates (CFPs) for neurosurgical bone flap fixation.

Materials And Methods: The study group comprised five patients who were intra-individually examined with 3 and 7 T MRI preoperatively and postoperatively (within 72 h/3 months) after implantation of CFPs. Acquired sequences included T1-weighted magnetization-prepared rapid-acquisition gradient-echo (MPRAGE), T2-weighted turbo-spin-echo (TSE) imaging, and susceptibility-weighted imaging (SWI). Two experienced neurosurgeons and a neuroradiologist rated image quality and the presence of artifacts in consensus reading.

Results: Minor artifacts occurred around the CFPs in MPRAGE and T2 TSE at both field strengths, with no significant differences between 3 and 7 T. In SWI, artifacts were accentuated in the early postoperative scans at both field strengths due to intracranial air and hemorrhagic remnants. After resorption, the brain tissue directly adjacent to skull bone could still be assessed. Image quality after 3 months was equal to the preoperative examinations at 3 and 7 T.

Conclusion: Image quality after CFP implantation was not significantly impaired in 7 T MRI, and artifacts were comparable to those in 3 T MRI.

Citing Articles

Central intra-lesional iron deposits as a possible novel imaging marker at 7 Tesla MRI in Susac Syndrome - an exploratory study.

Strunk D, Sinnecker T, Kleffner I, Doerr J, Ringelstein M, Gross C BMC Med Imaging. 2024; 24(1):4.

PMID: 38166655 PMC: 10759674. DOI: 10.1186/s12880-023-01171-7.


Safety and function of programmable ventriculo-peritoneal shunt valves: An in vitro 7 Tesla magnetic resonance imaging study.

Chen B, Dammann P, Jabbarli R, Sure U, Quick H, Kraff O PLoS One. 2023; 18(10):e0292666.

PMID: 37819939 PMC: 10566673. DOI: 10.1371/journal.pone.0292666.


Use of a Commercial 7-T MRI Scanner for Clinical Brain Imaging: Indications, Protocols, Challenges, and Solutions-A Single-Center Experience.

Ozutemiz C, White M, Elvendahl W, Eryaman Y, Marjanska M, Metzger G AJR Am J Roentgenol. 2023; 221(6):788-804.

PMID: 37377363 PMC: 10825876. DOI: 10.2214/AJR.23.29342.


Development and evaluation of a numerical simulation approach to predict metal artifacts from passive implants in MRI.

Spronk T, Kraff O, Kreutner J, Schaefers G, Quick H MAGMA. 2021; 35(3):485-497.

PMID: 34655346 PMC: 9188622. DOI: 10.1007/s10334-021-00966-5.


7T MR Safety.

Fagan A, Bitz A, Bjorkman-Burtscher I, Collins C, Kimbrell V, Raaijmakers A J Magn Reson Imaging. 2020; 53(2):333-346.

PMID: 32830900 PMC: 8170917. DOI: 10.1002/jmri.27319.


References
1.
Kalpathy-Cramer J, Gerstner E, Emblem K, Andronesi O, Rosen B . Advanced magnetic resonance imaging of the physical processes in human glioblastoma. Cancer Res. 2014; 74(17):4622-4637. PMC: 4155518. DOI: 10.1158/0008-5472.CAN-14-0383. View

2.
Grabner G, Nobauer I, Elandt K, Kronnerwetter C, Woehrer A, Marosi C . Longitudinal brain imaging of five malignant glioma patients treated with bevacizumab using susceptibility-weighted magnetic resonance imaging at 7 T. Magn Reson Imaging. 2011; 30(1):139-47. DOI: 10.1016/j.mri.2011.08.004. View

3.
Feng D, McCauley J, Morgan-Curtis F, Salam R, Pennell D, Loveless M . Evaluation of 39 medical implants at 7.0 T. Br J Radiol. 2015; 88(1056):20150633. PMC: 4984944. DOI: 10.1259/bjr.20150633. View

4.
Wrede K, Dammann P, Johst S, Monninghoff C, Schlamann M, Maderwald S . Non-Enhanced MR Imaging of Cerebral Arteriovenous Malformations at 7 Tesla. Eur Radiol. 2015; 26(3):829-39. DOI: 10.1007/s00330-015-3875-0. View

5.
Ieva A, God S, Grabner G, Grizzi F, Sherif C, Matula C . Three-dimensional susceptibility-weighted imaging at 7 T using fractal-based quantitative analysis to grade gliomas. Neuroradiology. 2012; 55(1):35-40. DOI: 10.1007/s00234-012-1081-1. View