» Articles » PMID: 26682167

Comparison of 45-degree Kelman and 45-degree Balanced Phaco Tip Designs in Torsional Microcoaxial Phacoemulsification

Overview
Specialty Ophthalmology
Date 2015 Dec 19
PMID 26682167
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To compare the intraoperative performance and postoperative outcome after microcoaxial torsional phacoemulsification using either a Kelman or balanced phaco tip.

Methods: Cataracts were treated using 2.2 mm microcoaxial torsional phacoemulsification using either a 45-degree mini-flared Kelman(®) or a 45-degree Intrepid(®) Balanced phaco tip. Intraoperative measurements included total ultrasound (US) time, cumulative dissipated energy (CDE), torsional US time, and balanced salt solution (BSS) use. The central endothelial cell density (ECD) and central corneal thickness (CCT) were evaluated preoperatively and postoperatively 1, 7, and 30d after surgery using noncontact specular microscopy.

Results: The 116 enrolled eyes (116 patients) were divided equally between the Kelman and balanced tip groups. Intraoperative measurements showed significantly less total US time, torsional US time, CDE, and BSS use in the balanced group than in Kelman group (P<0.05). The total US time, torsional US time, CDE, and BSS use were 17.45±14.53s, 16.63±13.97s, 6.38±5.26, and 48.21±17.21 mL in the Kelman group and 11.39 ± 9.60s, 10.90 ± 9.25s, 4.04 ± 3.42, and 41.36 ± 12.70 mL in the balanced group, respectively.

Conclusion: Torsional phacoemulsification performed with a balanced tip provided more effective lens removal with less total US time, torsional time, CDE, and BSS use, as well as similar changes in ECD with a Kelman tip in all cataract grades. This special designed phaco tip for torsional phacoemulsification provides an alternative phaco tip for many surgeons' preference with straight phaco tip.

Citing Articles

Comparison of Phacoemulsification and Aspiration Parameters in Cataract Surgery: Metal Tip vs. Hybrid Tip.

Ichikawa K, Ichikawa K, Tokiwa S, Sato Y, Miyazaki T, Tanaka Y Bioengineering (Basel). 2025; 11(12.

PMID: 39768013 PMC: 11727584. DOI: 10.3390/bioengineering11121195.


The Cutting Efficiency of a Hybrid Phacoemulsification Tip Using High and Low Intraocular Pressure Settings in Different Grades of Cataract.

Spaulding J, Hall B Clin Ophthalmol. 2024; 18:3271-3276.

PMID: 39555213 PMC: 11569703. DOI: 10.2147/OPTH.S488581.


Comparative Clinical Study of Surgical Performance of Quatera 700 versus Centurion and Signature Pro Phacoemulsification Systems.

Ganesh S, Brar S, Sriganesh S, Bhavsar H Clin Ophthalmol. 2024; 18:2685-2695.

PMID: 39346969 PMC: 11430833. DOI: 10.2147/OPTH.S456660.


In vitro analysis of clinically relevant aspects of a polymer-coated phacoemulsification tip vs a traditional tip.

Zacharias J J Cataract Refract Surg. 2023; 49(12):1264-1269.

PMID: 37464557 PMC: 10664810. DOI: 10.1097/j.jcrs.0000000000001267.


Cumulative dissipated energy in eyes with and without corneal opacity.

Ilhan C, Kocak Altintas A Rom J Ophthalmol. 2022; 66(3):233-239.

PMID: 36349172 PMC: 9585489. DOI: 10.22336/rjo.2022.45.


References
1.
Kim J, Chung S, Joo C . Clinical application of a Scheimpflug system for lens density measurements in phacoemulsification. J Cataract Refract Surg. 2009; 35(7):1204-9. DOI: 10.1016/j.jcrs.2009.02.032. View

2.
Demircan S, Gokce G, Atas M, Baskan B, Goktas E, Zararsiz G . The Impact of Reused Phaco Tip on Outcomes of Phacoemulsification Surgery. Curr Eye Res. 2015; 41(5):636-42. DOI: 10.3109/02713683.2015.1039654. View

3.
Watanabe A . New phacoemulsification tip with a grooved, threaded-tip construction. J Cataract Refract Surg. 2011; 37(7):1329-32. DOI: 10.1016/j.jcrs.2011.01.030. View

4.
Helvacioglu F, Yeter C, Sencan S, Tunc Z, Uyar O . Comparison of two different ultrasound methods of phacoemulsification. Am J Ophthalmol. 2014; 158(2):221-226.e1. DOI: 10.1016/j.ajo.2014.04.015. View

5.
Atas M, Demircan S, Karatepe Hashas A, Gulhan A, Zararsiz G . Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines. Int J Ophthalmol. 2014; 7(5):822-7. PMC: 4206888. DOI: 10.3980/j.issn.2222-3959.2014.05.15. View