Two-piece Zirconia Oral Implants Withstand Masticatory Loads: An Investigation in the Artificial Mouth

Overview
Publisher Elsevier
Date 2015 Aug 24
PMID 26298800
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate the fracture resistance of two-piece zirconia oral implants after long-term thermomechanical cycling in an aqueous environment. Non-loaded samples and a one-piece implant system served as control groups.

Methods: A total of 48 zirconia implants were evaluated: 16 one-piece implants (ATZ; Group A) and 32 differently connected two-piece implants (16 screwed, Group B; 16 bonded, Group C) made of Y-TZP-A (implant+abutment; B) and Y-TZP-A/ATZ (implant/abutment; C), respectively. These groups were divided into two subgroups composed of 8 samples. The samples of subgroups 1 (A1, B1, C1) were not exposed to any cyclic loading, whereas subgroups 2 (A2, B2, C2) were loaded with 10 million cycles (98 N). Subsequently, all 48 implants were statically loaded to fracture.

Results: A constant load on distinct lever arms resulted in different exerted bending moments during the dynamic loading (A2: 23.4Ncm, B2: 17.9 Ncm, C2: 32.3 Ncm). All implants survived the long-term thermomechanical cycling. For the static loading the following average bending moments were calculated: A1/A2: 362/399 Ncm; B1/B2: 398/346 Ncm; C1/C2: 380/252 Ncm. Foregoing dynamic loading significantly increased fracture resistance of Group A implants, whereas Group B/C implants showed significantly decreased values. Potentially owed to the experimental setup in an aqueous environment of 60 °C, 5/8 C2 samples showed mobility between implant and abutment due to debonding after dynamic loading conditions.

Significance: The evaluated ceramic implant systems seem to be able to resist physiological chewing forces long-term. Within the limitations of the experimental setup, the connecting mechanism of Group C implants might be a weak point.

Citing Articles

The Fracture Resistance Comparison between Titanium and Zirconia Implant Abutments with and without Ageing: Systematic Review and Meta-Analysis.

Chmielewski M, Dabrowski W, Ordyniec-Kwasnica I Dent J (Basel). 2024; 12(9).

PMID: 39329840 PMC: 11431843. DOI: 10.3390/dj12090274.


Clinical and radiographic outcomes of a two-piece ceramic implant: one year results from a prospective clinical trial.

da Silva A, Horta Dos Santos F, Mota R, Teixeira M, Telles D, Lourenco E Clin Oral Investig. 2024; 28(7):380.

PMID: 38886209 DOI: 10.1007/s00784-024-05783-3.


Finite Element Analysis and Fatigue Test of INTEGRA Dental Implant System.

Zielinski R, Lipa S, Piechaczek M, Sowinski J, Kolkowska A, Simka W Materials (Basel). 2024; 17(5).

PMID: 38473684 PMC: 10934518. DOI: 10.3390/ma17051213.


Fracture Resistance of a Two-Piece Zirconia Implant System after Artificial Loading and/or Hydrothermal Aging-An In Vitro Investigation.

Kohal R, Schikofski T, Adolfsson E, Vach K, Patzelt S, Nold J J Funct Biomater. 2023; 14(12).

PMID: 38132821 PMC: 10743638. DOI: 10.3390/jfb14120567.


Long-Term Stability of Hydrothermally Aged and/or Dynamically Loaded One-Piece Diameter Reduced Zirconia Oral Implants.

Kohal R, Trinkner A, Burkhardt F, Patzelt S, Vach K, Kuster M J Funct Biomater. 2023; 14(3).

PMID: 36976047 PMC: 10051462. DOI: 10.3390/jfb14030123.