» Articles » PMID: 26257847

The Changes of Non-invasive Hemoglobin and Perfusion Index of Pulse CO-Oximetry During Induction of General Anesthesia

Overview
Specialty Anesthesiology
Date 2015 Aug 11
PMID 26257847
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background: We hypothesized that induction of general anesthesia using sevoflurane improves the accuracy of non-invasive hemoglobin (SpHb) measurement of Masimo Radical-7® Pulse CO-Oximetry by inducing peripheral vasodilation and increasing the perfusion index (PI). The aim of this study is to investigate the change in the SpHb and the PI measured by Rad7 during induction of general anesthesia using sevoflurane.

Methods: The laboratory hemoglobin (Hblab) was measured before surgery by venous blood sampling. The SpHb and the PI was measured twice; before and after the induction of general anesthesia using sevoflurane. The changes of SpHb, Hbbias (Hbbias = SpHb - Hblab), and PI before and after the induction of general anesthesia were analyzed using a paired t-test. Also, a Pearson correlation coefficient analysis was used to analyze the correlation between the Hbbias and the PI.

Results: The SpHb and the PI were increased after the induction of general anesthesia using sevoflurane. There was a statistically significant change in the Hbbias from -2.8 to -0.7 after the induction of general anesthesia. However, the limit of agreement (2 SD) of the Hbbias did not change after the induction of general anesthesia. The Pearson correlation coefficient between the Hbbias and the PI was not statistically significant.

Conclusions: During induction of general anesthesia using sevoflurane, the accuracy of SpHb measurement was improved and precision was not changed. The correlation between Hbbias and PI was not significant.

Citing Articles

Accuracy of continuous and real-time total hemoglobin during bimaxillary orthognathic surgery.

Kim C, Yoon J, Yun G, Kim H, Kim E J Dent Anesth Pain Med. 2025; 25(1):23-32.

PMID: 39944844 PMC: 11811515. DOI: 10.17245/jdapm.2025.25.1.23.


Comparing Non-Invasive Spectrophotometry to Hematology Analysis for Hemoglobin Measurements in Sickle Cell Disease Patients.

Yassen K, Omer N, Alsahaf F, Al Amer F, Alhamad F, Alherz I J Clin Med. 2023; 12(24).

PMID: 38137588 PMC: 10744205. DOI: 10.3390/jcm12247517.


Can Non-Invasive Spectrophotometric Hemoglobin Replace Laboratory Hemoglobin Concentrations for Preoperative Anemia Screening? A Diagnostic Test Accuracy Study.

Alwabari M, Alhamad F, Alsahaf F, Al Amer F, Alniniya F, Alherz I J Clin Med. 2023; 12(17).

PMID: 37685800 PMC: 10488634. DOI: 10.3390/jcm12175733.


Evaluation of prediction effect of perfusion index for supraclavicular brachial plexus block in children: protocol for a randomized trial.

Wang J, Deng L, Xu A Trials. 2022; 23(1):629.

PMID: 35927745 PMC: 9351202. DOI: 10.1186/s13063-022-06597-y.


Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements.

Spronk A, De Jong R, Koopman S, Hoeks S, Stolker R Anaesthesiol Intensive Ther. 2021; 53(2):103-107.

PMID: 34006050 PMC: 10158438. DOI: 10.5114/ait.2021.105658.


References
1.
Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus J, Mathieu E . Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011; 57(4):330-3. DOI: 10.1016/j.annemergmed.2010.11.032. View

2.
Berkow L, Rotolo S, Mirski E . Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011; 113(6):1396-402. DOI: 10.1213/ANE.0b013e318230b425. View

3.
Nguyen B, Vincent J, Nowak E, Coat M, Paleiron N, Gouny P . The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011; 113(5):1052-7. DOI: 10.1213/ANE.0b013e31822c9679. View

4.
Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B . Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology. 2011; 115(3):548-54. DOI: 10.1097/ALN.0b013e3182270c22. View

5.
Albertin A, La Colla L, Gandolfi A, Colnaghi E, Mandelli D, Gioia G . Greater peripheral blood flow but less bleeding with propofol versus sevoflurane during spine surgery: a possible physiologic model?. Spine (Phila Pa 1976). 2008; 33(18):2017-22. DOI: 10.1097/BRS.0b013e31817e0405. View