A Comparison of Three Methods of Hemoglobin Monitoring in Patients Undergoing Spine Surgery

Overview
Journal Anesth Analg
Specialty Anesthesiology
Date 2011 Mar 10
PMID 21385985
Citations 39
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Hemoglobin values (Hb) can facilitate decisions regarding perioperative transfusion management. Currently, Hb can be determined invasively by analyzing blood via laboratory Co-Oximetry (tHb) or by point-of-care HemoCue (HCue). Recently, a new noninvasive, continuous spectrophotometric sensor (Masimo SpHb) was introduced into clinical practice. We compared the accuracy of the SpHb and HCue with tHb.

Methods: Twenty patients, ages 40 to 80 years, were studied. They received general anesthesia and underwent spine surgery in the prone position. All blood samples were obtained from a radial artery catheter. SpHb, tHb, and HCue were determined immediately after induction of anesthesia, but before the start of surgery and approximately every hour thereafter. Primary outcomes were defined on the basis of the following differences between measures: SpHb - tHb or HCue - tHb. All patients had 3 to 5 observations taken on each measure. Differences and absolute differences were analyzed by several techniques to assess accuracy. We also investigated the relationship between observed differences and the following variables: tHb level, duration of surgery, age, weight, and perfusion index.

Results: Data consisted of 78 measurements of SpHb, tHb, and HCue made on the 20 patients. Absolute differences between SpHb and tHb were <1.5 g/dL for 61% of observations, between 1.6 to 2.0 g/dL for 16% and >2.0 g/dL for 22% of the observations. Observed differences displayed significant decreases with time and higher perfusion index values. No systematic relationships were observed with age or weight. Except for 1 value, all of the HCue values were <1.0 g/dL of tHb.

Conclusions: Although HCue was consistently accurate, our data confirm that SpHb often correlated well with tHb values. Yet our study indicates that SpHb may not be as accurate as clinically necessary in some patients. Improved refinement of continuous, noninvasive technology, such as SpHb, could address important clinical requirements.

Citing Articles

Accuracy of continuous and real-time total hemoglobin during bimaxillary orthognathic surgery.

Kim C, Yoon J, Yun G, Kim H, Kim E J Dent Anesth Pain Med. 2025; 25(1):23-32.

PMID: 39944844 PMC: 11811515. DOI: 10.17245/jdapm.2025.25.1.23.


Effect of hyperbilirubinemia on the accuracy of continuous non-invasive hemoglobin measurements in liver transplantation recipients.

Yoon S, Jung C, Kim T, Lee H Sci Rep. 2024; 14(1):5072.

PMID: 38429444 PMC: 10907682. DOI: 10.1038/s41598-024-55837-5.


Accuracy of point-of-care testing devices for haemoglobin in the operating room: meta-analysis.

Ahn H, Lenet T, Gilbert R, Mallick R, Shaw J, Fergusson D BJS Open. 2024; 8(1).

PMID: 38266123 PMC: 10807999. DOI: 10.1093/bjsopen/zrad148.


Comparing Non-Invasive Spectrophotometry to Hematology Analysis for Hemoglobin Measurements in Sickle Cell Disease Patients.

Yassen K, Omer N, Alsahaf F, Al Amer F, Alhamad F, Alherz I J Clin Med. 2023; 12(24).

PMID: 38137588 PMC: 10744205. DOI: 10.3390/jcm12247517.


Can Non-Invasive Spectrophotometric Hemoglobin Replace Laboratory Hemoglobin Concentrations for Preoperative Anemia Screening? A Diagnostic Test Accuracy Study.

Alwabari M, Alhamad F, Alsahaf F, Al Amer F, Alniniya F, Alherz I J Clin Med. 2023; 12(17).

PMID: 37685800 PMC: 10488634. DOI: 10.3390/jcm12175733.