Performance Evaluation of a Noninvasive Hemoglobin Monitoring Device

Overview
Journal Ann Emerg Med
Specialty Emergency Medicine
Date 2011 Jan 22
PMID 21251729
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Study Objective: Hemoglobin measurement is a routine procedure, and a noninvasive point-of-care device may increase the quality of care. The aim of the present study is to compare hemoglobin concentration values obtained with a portable totally noninvasive device, the Masimo Labs Radical-7 Pulse CO-Oximeter, with the results obtained by the ADVIA 2120 in the laboratory.

Methods: This was a prospective monocentric open trial enrolling patients consulting in the emergency department of a university hospital from June 16 to December 17, 2009. The main outcome measure was the agreement between both methods and evaluation of the percentage of potential decision error for transfusion.

Results: Samples from 300 consecutive patients were assessed. Hemoglobin concentration could not be obtained with the new device for 24 patients. In others, the mean bias, the lower and the upper limits of agreement between the 2 methods, was 1.8 g/dL (95% estimated confidence interval [CI] 1.5 to 2.1 g/dL), -3.3 g/dL (95% CI -3.8 to -2.8 g/dL), and 6.9 g/dL (95% CI 6.4 to 7.4 g/dL), respectively. The intraclass correlation coefficient was 0.53 (estimated 95% CI 0.10 to 0.74). The number of potential errors about transfusion decision was 38 (13% of patients). The peripheral oxygen saturation and the true value of hemoglobin concentration were independently associated with the bias.

Conclusion: Results from this widely available noninvasive point-of-care hemoglobin monitoring device were systematically biased and too unreliable to guide transfusion decisions.

Citing Articles

Smartphone-based point-of-care anemia screening in rural Bihar in India.

Haggenmuller V, Bogler L, Weber A, Kumar A, Barnighausen T, Danquah I Commun Med (Lond). 2023; 3(1):38.

PMID: 36949164 PMC: 10033918. DOI: 10.1038/s43856-023-00267-z.


Hemoglobin Determination Using Pulse Co-Oximetry and Reduced-Volume Blood Gas Analysis in the Critically Ill: A Prospective Cohort Study.

Czempik P, Pluta M, Krzych L Diagnostics (Basel). 2022; 12(12).

PMID: 36552914 PMC: 9776962. DOI: 10.3390/diagnostics12122908.


Continuous hemoglobin measurement during frontal advancement operations can improve patient outcomes.

Saracoglu A, Abdullayev R, Sakar M, Sacak B, Girgin Incekoy F, Aykac Z J Clin Monit Comput. 2022; 36(6):1689-1695.

PMID: 35257263 DOI: 10.1007/s10877-022-00813-5.


Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Shabaninejad H, Ghadimi N, Sayehmiri K, Hosseinifard H, Azarfarin R, Gorji H J Anesth. 2019; 33(3):441-453.

PMID: 30895376 DOI: 10.1007/s00540-019-02629-1.


Non-invasive detection of severe neutropenia in chemotherapy patients by optical imaging of nailfold microcirculation.

Bourquard A, Pablo-Trinidad A, Butterworth I, Sanchez-Ferro A, Cerrato C, Humala K Sci Rep. 2018; 8(1):5301.

PMID: 29593221 PMC: 5871877. DOI: 10.1038/s41598-018-23591-0.