Evaluation of an Endosseous Oral Implant System According to UNI EN ISO 14801 Fatigue Test Protocol

Overview
Journal Implant Dent
Date 2014 Oct 8
PMID 25290279
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aim of this in vitro study was to evaluate the maximum fracture load and the mechanical performance to cyclic fatigue stress of a dental implant system.

Methods: An experimental study according to the international standards (UNI EN ISO 14801: 2008) was carried out using 15 implants (3.80 mm of diameter and 13 mm of length) with applied straight titanium abutment tightened to 30 N. Five samples were subjected to compression stress at break. Based on the mean fracture load value obtained in this test, the levels of dynamic loading range were set that were carried on at a frequency of 15 Hz for 5 × 10 cycles.

Results: The compression stress at break mean value of the tested implants was 499.40 N (SD ±50.1 N). In the mechanical fatigue stress test, the fatigue limit for 5 × 10 load cycles for all tested samples was 250 N.

Conclusions: The evaluated implant system proved to withstand considerable mechanical loads under the "worst-case" loading situation performed according to UNI EN ISO 14801 standard. The reliability of this test protocol makes it suitable to be accomplished for understanding and comparing mechanical properties of other implant systems.

Citing Articles

Influence of Narrow Titanium Dental Implant Diameter on Fatigue Behavior: A Comparison between Unitary and Splinted Implants.

Gonzalez Terrats R, Linares M, Punset M, Molmeneu M, Nart Molina J, Ruiz Magaz V J Clin Med. 2024; 13(6).

PMID: 38541858 PMC: 10971635. DOI: 10.3390/jcm13061632.


Finite Element Analysis and Fatigue Test of INTEGRA Dental Implant System.

Zielinski R, Lipa S, Piechaczek M, Sowinski J, Kolkowska A, Simka W Materials (Basel). 2024; 17(5).

PMID: 38473684 PMC: 10934518. DOI: 10.3390/ma17051213.


Implant-Supported Fixed Partial Dentures with Posterior Cantilevers: In Vitro Study of Mechanical Behavior.

Garcia-Sala Bonmati F, Bustamante-Hernandez N, Alonso Perez-Barquero J, Maneiro-Lojo J, Fons-Badal C, Labaig-Caturla C Materials (Basel). 2023; 16(20).

PMID: 37895786 PMC: 10608170. DOI: 10.3390/ma16206805.


Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees.

Tolesano Loureiro K, Moura M, de Oliveira G, Novais V, Zancope K, Simamoto Junior P Healthcare (Basel). 2023; 11(12).

PMID: 37372893 PMC: 10298607. DOI: 10.3390/healthcare11121775.


Biological Oriented Immediate Loading: A New Mathematical Implant Vertical Insertion Protocol, Five-Year Follow-Up Study.

Bambini F, Orilisi G, Quaranta A, Meme L Materials (Basel). 2021; 14(2).

PMID: 33466862 PMC: 7830575. DOI: 10.3390/ma14020387.