Abstract

Objectives: Aim of the study was to assess the feasibility and to compare three non-enhanced T1-weighted (w) sequences for liver vessel imaging at 7 Tesla (T).

Material And Methods: 12 healthy volunteers were examined on a 7 T whole-body MR-system. The following non-enhanced sequences were acquired: T1w 2D FLASH, T1w 3D FLASH and Time of flight (TOF)-MRA. Qualitative image analysis was performed by two radiologists including over all image quality as well as vessel delineation of the liver arteries, liver veins and portal vein and the presence of artifacts using a five-point scale (5 = excellent vessel delineation to 1 = non-diagnostic). Contrast ratios (CR), SNR und CNR of the above named vessels in correlation to adjacent liver tissue were calculated for quantitative assessment. For statistical analysis, a Wilcoxon Rank Test was applied.

Results: All three sequences provided a homogenous hyperintense delineation of the assessed liver vessels. Qualitative image analysis demonstrated the superiority of TOF-MRA, providing best overall image quality (TOF 4.17, 2D FLASH 3.42, 3D FLASH 3.46; p<0.01) as well as highest image quality values for all analyzed liver vessel segments. TOF-MRA was least impaired by B1 inhomogeneity (4.13) and susceptibility artifacts (4.63) out of all three sequences (p<0.01). Quantitative image analysis confirmed the superiority of TOF MRA showing significant higher CR values for all liver vessels (e.g. right hepatic artery TOF 0.47, 2D FLASH 0.09, 3D FLASH 0.11 with p = 0.02 and 0.01, respectively). Providing the lowest standard deviation in noise, TOF showed highest values for SNR and CNR.

Conclusions: Non-enhanced T1w imaging in general and TOF MRA in particular, appear to be promising techniques for high quality non-enhanced liver vessel assessment at 7 T.

Citing Articles

From strength to precision: A systematic review exploring the clinical utility of 7-Tesla magnetic resonance imaging in abdominal imaging.

Perera Molligoda Arachchige A, Teixeira de Castro Goncalves Ortega A, Catapano F, Politi L, Hoff M World J Radiol. 2024; 16(1):20-31.

PMID: 38312348 PMC: 10835428. DOI: 10.4329/wjr.v16.i1.20.


Evolution of UHF Body Imaging in the Human Torso at 7T: Technology, Applications, and Future Directions.

Erturk M, Li X, Van de Moortele P, Ugurbil K, Metzger G Top Magn Reson Imaging. 2019; 28(3):101-124.

PMID: 31188271 PMC: 6587233. DOI: 10.1097/RMR.0000000000000202.


Parallel transmission for ultrahigh-field imaging.

Padormo F, Beqiri A, Hajnal J, Malik S NMR Biomed. 2015; 29(9):1145-61.

PMID: 25989904 PMC: 4995736. DOI: 10.1002/nbm.3313.

References
1.
Sahani D, Mehta A, Blake M, Prasad S, Harris G, Saini S . Preoperative hepatic vascular evaluation with CT and MR angiography: implications for surgery. Radiographics. 2004; 24(5):1367-80. DOI: 10.1148/rg.245035224. View

2.
Duan Q, van Gelderen P, Duyn J . Improved Bloch-Siegert based B1 mapping by reducing off-resonance shift. NMR Biomed. 2013; 26(9):1070-8. PMC: 3669656. DOI: 10.1002/nbm.2920. View

3.
Grobner T . Gadolinium--a specific trigger for the development of nephrogenic fibrosing dermopathy and nephrogenic systemic fibrosis?. Nephrol Dial Transplant. 2006; 21(4):1104-8. DOI: 10.1093/ndt/gfk062. View

4.
Lee M, Lee J, Lee J, Kim S, Park E, Han J . Preoperative evaluation of hepatic arterial and portal venous anatomy using the time resolved echo-shared MR angiographic technique in living liver donors. Eur Radiol. 2006; 17(4):1074-80. DOI: 10.1007/s00330-006-0447-3. View

5.
Uchida K, Taniguchi M, Shimamura T, Suzuki T, Yamashita K, Ota M . Three-dimensional computed tomography scan analysis of hepatic vasculatures in the donor liver for living donor liver transplantation. Liver Transpl. 2010; 16(9):1062-8. DOI: 10.1002/lt.22109. View