ECG-based Gating in Ultra High Field Cardiovascular Magnetic Resonance Using an Independent Component Analysis Approach

Overview
Publisher Elsevier
Date 2013 Nov 21
PMID 24252594
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Background: In Cardiovascular Magnetic Resonance (CMR), the synchronization of image acquisition with heart motion is performed in clinical practice by processing the electrocardiogram (ECG). The ECG-based synchronization is well established for MR scanners with magnetic fields up to 3 T. However, this technique is prone to errors in ultra high field environments, e.g. in 7 T MR scanners as used in research applications. The high magnetic fields cause severe magnetohydrodynamic (MHD) effects which disturb the ECG signal. Image synchronization is thus less reliable and yields artefacts in CMR images.

Methods: A strategy based on Independent Component Analysis (ICA) was pursued in this work to enhance the ECG contribution and attenuate the MHD effect. ICA was applied to 12-lead ECG signals recorded inside a 7 T MR scanner. An automatic source identification procedure was proposed to identify an independent component (IC) dominated by the ECG signal. The identified IC was then used for detecting the R-peaks. The presented ICA-based method was compared to other R-peak detection methods using 1) the raw ECG signal, 2) the raw vectorcardiogram (VCG), 3) the state-of-the-art gating technique based on the VCG, 4) an updated version of the VCG-based approach and 5) the ICA of the VCG.

Results: ECG signals from eight volunteers were recorded inside the MR scanner. Recordings with an overall length of 87 min accounting for 5457 QRS complexes were available for the analysis. The records were divided into a training and a test dataset. In terms of R-peak detection within the test dataset, the proposed ICA-based algorithm achieved a detection performance with an average sensitivity (Se) of 99.2%, a positive predictive value (+P) of 99.1%, with an average trigger delay and jitter of 5.8 ms and 5.0 ms, respectively. Long term stability of the demixing matrix was shown based on two measurements of the same subject, each being separated by one year, whereas an averaged detection performance of Se = 99.4% and +P = 99.7% was achieved.Compared to the state-of-the-art VCG-based gating technique at 7 T, the proposed method increased the sensitivity and positive predictive value within the test dataset by 27.1% and 42.7%, respectively.

Conclusions: The presented ICA-based method allows the estimation and identification of an IC dominated by the ECG signal. R-peak detection based on this IC outperforms the state-of-the-art VCG-based technique in a 7 T MR scanner environment.

Citing Articles

Comprehensive review of artifacts in cardiac MRI and their mitigation.

Rafiee M, Eyre K, Leo M, Benovoy M, Friedrich M, Chetrit M Int J Cardiovasc Imaging. 2024; 40(10):2021-2039.

PMID: 39292396 DOI: 10.1007/s10554-024-03234-4.


Verifying the Accuracy of Hemodynamic Analysis Using High Spatial Resolution 3D Phase-contrast MR Imaging on a 7T MR System: Comparison with a 3T System.

Tajima S, Isoda H, Fukunaga M, Komori Y, Naganawa S, Sadato N Magn Reson Med Sci. 2023; 24(1):88-102.

PMID: 38123345 PMC: 11733508. DOI: 10.2463/mrms.mp.2023-0016.


Cardiac magnetic resonance of hypertrophic heart phenotype: A review.

Tore D, Faletti R, Gaetani C, Bozzo E, Biondo A, Carisio A Heliyon. 2023; 9(6):e17336.

PMID: 37441401 PMC: 10333467. DOI: 10.1016/j.heliyon.2023.e17336.


Self-Attention MHDNet: A Novel Deep Learning Model for the Detection of R-Peaks in the Electrocardiogram Signals Corrupted with Magnetohydrodynamic Effect.

Chowdhury M, Chowdhury M, Khan M, Ullah M, Mahmud S, Khandakar A Bioengineering (Basel). 2023; 10(5).

PMID: 37237612 PMC: 10215845. DOI: 10.3390/bioengineering10050542.


Free-breathing, non-ECG, simultaneous myocardial T , T , T *, and fat-fraction mapping with motion-resolved cardiovascular MR multitasking.

Cao T, Wang N, Kwan A, Lee H, Mao X, Xie Y Magn Reson Med. 2022; 88(4):1748-1763.

PMID: 35713184 PMC: 9339519. DOI: 10.1002/mrm.29351.


References
1.
Scott A, Keegan J, Firmin D . Motion in cardiovascular MR imaging. Radiology. 2009; 250(2):331-51. DOI: 10.1148/radiol.2502071998. View

2.
Kyriakou A, Neufeld E, Szczerba D, Kainz W, Luechinger R, Kozerke S . Patient-specific simulations and measurements of the magneto-hemodynamic effect in human primary vessels. Physiol Meas. 2012; 33(2):117-30. DOI: 10.1088/0967-3334/33/2/117. View

3.
Felblinger J, Slotboom J, Kreis R, Jung B, Boesch C . Restoration of electrophysiological signals distorted by inductive effects of magnetic field gradients during MR sequences. Magn Reson Med. 1999; 41(4):715-21. DOI: 10.1002/(sici)1522-2594(199904)41:4<715::aid-mrm9>3.0.co;2-7. View

4.
Kinouchi Y, Yamaguchi H, Tenforde T . Theoretical analysis of magnetic field interactions with aortic blood flow. Bioelectromagnetics. 1996; 17(1):21-32. DOI: 10.1002/(SICI)1521-186X(1996)17:1<21::AID-BEM3>3.0.CO;2-8. View

5.
Frauenrath T, Hezel F, Renz W, de Geyer dOrth T, Dieringer M, von Knobelsdorff-Brenkenhoff F . Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla. J Cardiovasc Magn Reson. 2010; 12:67. PMC: 2998500. DOI: 10.1186/1532-429X-12-67. View