Self-Attention MHDNet: A Novel Deep Learning Model for the Detection of R-Peaks in the Electrocardiogram Signals Corrupted with Magnetohydrodynamic Effect

Overview
Date 2023 May 27
PMID 37237612
Authors
Affiliations
Soon will be listed here.
Abstract

Magnetic resonance imaging (MRI) is commonly used in medical diagnosis and minimally invasive image-guided operations. During an MRI scan, the patient's electrocardiogram (ECG) may be required for either gating or patient monitoring. However, the challenging environment of an MRI scanner, with its several types of magnetic fields, creates significant distortions of the collected ECG data due to the Magnetohydrodynamic (MHD) effect. These changes can be seen as irregular heartbeats. These distortions and abnormalities hamper the detection of QRS complexes, and a more in-depth diagnosis based on the ECG. This study aims to reliably detect R-peaks in the ECG waveforms in 3 Tesla (T) and 7T magnetic fields. A novel model, Self-Attention MHDNet, is proposed to detect R peaks from the MHD corrupted ECG signal through 1D-segmentation. The proposed model achieves a recall and precision of 99.83% and 99.68%, respectively, for the ECG data acquired in a 3T setting, while 99.87% and 99.78%, respectively, in a 7T setting. This model can thus be used in accurately gating the trigger pulse for the cardiovascular functional MRI.

Citing Articles

Verifying the Accuracy of Hemodynamic Analysis Using High Spatial Resolution 3D Phase-contrast MR Imaging on a 7T MR System: Comparison with a 3T System.

Tajima S, Isoda H, Fukunaga M, Komori Y, Naganawa S, Sadato N Magn Reson Med Sci. 2023; 24(1):88-102.

PMID: 38123345 PMC: 11733508. DOI: 10.2463/mrms.mp.2023-0016.

References
1.
Kiranyaz S, Malik J, Abdallah H, Ince T, Iosifidis A, Gabbouj M . Self-organized Operational Neural Networks with Generative Neurons. Neural Netw. 2021; 140:294-308. DOI: 10.1016/j.neunet.2021.02.028. View

2.
Hamilton-Craig C, Staeb D, Al Najjar A, OBrien K, Crawford W, Fletcher S . 7-Tesla Functional Cardiovascular MR Using Vectorcardiographic Triggering-Overcoming the Magnetohydrodynamic Effect. Tomography. 2021; 7(3):323-332. PMC: 8396263. DOI: 10.3390/tomography7030029. View

3.
Spicher N, Kukuk M, Maderwald S, Ladd M . Initial evaluation of prospective cardiac triggering using photoplethysmography signals recorded with a video camera compared to pulse oximetry and electrocardiography at 7T MRI. Biomed Eng Online. 2016; 15(1):126. PMC: 5122166. DOI: 10.1186/s12938-016-0245-3. View

4.
Abi-Abdallah D, Chauvet E, Bouchet-Fakri L, Bataillard A, Briguet A, Fokapu O . Reference signal extraction from corrupted ECG using wavelet decomposition for MRI sequence triggering: application to small animals. Biomed Eng Online. 2006; 5:11. PMC: 1409788. DOI: 10.1186/1475-925X-5-11. View

5.
Petersen S, Aung N, Sanghvi M, Zemrak F, Fung K, Paiva J . Reference ranges for cardiac structure and function using cardiovascular magnetic resonance (CMR) in Caucasians from the UK Biobank population cohort. J Cardiovasc Magn Reson. 2017; 19(1):18. PMC: 5304550. DOI: 10.1186/s12968-017-0327-9. View