Evaluation and Calibration of a Binocular Infrared Pupillometer for Measuring Relative Afferent Pupillary Defect

Overview
Specialties Neurology
Ophthalmology
Date 2012 Jan 17
PMID 22246058
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Binocular infrared pupillometry allows an estimate of the relative afferent pupillary defect (RAPD), designated the pupillometric RAPD (pRAPD). We calibrated the pRAPD of a commercially available pupillometer against neutral density filters (NDFs) of known attenuation. The performance of the pupillometer using its own proprietary algorithm is assessed and compared to that of alternative algorithms.

Methods: The pRAPDs of 50 healthy volunteers were measured with each of 4 filters of known attenuation: 0.0, 0.3, 0.6, and 0.9 log units, positioned unilaterally in the light stimulus pathway. The filter values were plotted against the pupillometer output, and the slope and intercept were used to determine a calibration factor. Corrected pRAPD results were used to assess physiological ranges of pRAPD. The sensitivity and specificity to 0.3 log unit differences between increasing filter densities using receiver operator characteristic (ROC) curves.

Results: The calibrated physiological pRAPD ranged from 0 to 0.22 log units. The area under the ROC curve for detecting unilateral simulated pRAPD of 0.3 log units, the simulated disease progression from 0.3 to 0.6 log units, and a further progression from 0.6 to 0.9 log units by NDFs was 0.99 (95% confidence interval [CI], 0.95-1.00), 0.86 (95% CI, 0.78-0.92), and 0.79 (95% CI, 0.70-0.87), respectively. The optimum discrimination was for detecting a unilateral simulated pRAPD of 0.3 log units; sensitivity and specificity was 98% (95% CI, 88%-99%).

Conclusion: The commercially available pupillometer detects the RAPD induced by the NDFs with high sensitivity and specificity. The results suggest that it is best for detecting unilateral early disease but potentially useful for assessing progression of disease.

Citing Articles

Diagnostic accuracy of a modularized, virtual-reality-based automated pupillometer for detection of relative afferent pupillary defect in unilateral optic neuropathies.

Negi R, Kalivemula M, Bisht K, Bhate M, Sachdeva V, Bharadwaj S Front Ophthalmol (Lausanne). 2024; 4:1396511.

PMID: 39290775 PMC: 11405164. DOI: 10.3389/fopht.2024.1396511.


A haploscope based binocular pupillometer system to quantify the dynamics of direct and consensual Pupillary Light Reflex.

Meethal N, Mazumdar D, Morshchavka S, Robben J, van der Steen J, George R Sci Rep. 2021; 11(1):21090.

PMID: 34702842 PMC: 8548319. DOI: 10.1038/s41598-021-00434-z.


Dilute pilocarpine test for diagnosis of Adie's tonic pupil.

Yoo Y, Hwang J, Yang H Sci Rep. 2021; 11(1):10089.

PMID: 33980910 PMC: 8115311. DOI: 10.1038/s41598-021-89148-w.


A 30 s test for quantitative assessment of a relative afferent pupillary defect (RAPD): the infrared pupillary asymmetry (IPA).

Meneguette N, de Carvalho J, Petzold A J Neurol. 2019; 266(4):969-974.

PMID: 30746557 DOI: 10.1007/s00415-019-09223-1.


Quantitative analysis of pupillometry in isolated third nerve palsy.

Kim H, Yang H, Hwang J PLoS One. 2018; 13(11):e0208259.

PMID: 30496292 PMC: 6264827. DOI: 10.1371/journal.pone.0208259.