» Articles » PMID: 22210571

A Randomized Clinical Trial to Assess the Efficacy and Safety of Real-time Continuous Glucose Monitoring in the Management of Type 1 Diabetes in Young Children Aged 4 to <10 Years

Overview
Journal Diabetes Care
Specialty Endocrinology
Date 2012 Jan 3
PMID 22210571
Citations 84
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Continuous glucose monitoring (CGM) has been demonstrated to improve glycemic control in adults with type 1 diabetes but less so in children. We designed a study to assess CGM benefit in young children aged 4 to 9 years with type 1 diabetes.

Research Design And Methods: After a run-in phase, 146 children with type 1 diabetes (mean age 7.5 ± 1.7 years, 64% on pumps, median diabetes duration 3.5 years) were randomly assigned to CGM or to usual care. The primary outcome was reduction in HbA(1c) at 26 weeks by ≥0.5% without the occurrence of severe hypoglycemia.

Results: The primary outcome was achieved by 19% in the CGM group and 28% in the control group (P = 0.17). Mean change in HbA(1c) was -0.1% in each group (P = 0.79). Severe hypoglycemia rates were similarly low in both groups. CGM wear decreased over time, with only 41% averaging at least 6 days/week at 26 weeks. There was no correlation between CGM use and change in HbA(1c) (r(s) = -0.09, P = 0.44). CGM wear was well tolerated, and parental satisfaction with CGM was high. However, parental fear of hypoglycemia was not reduced.

Conclusions: CGM in 4- to 9-year-olds did not improve glycemic control despite a high degree of parental satisfaction with CGM. We postulate that this finding may be related in part to limited use of the CGM glucose data in day-to-day management and to an unremitting fear of hypoglycemia. Overcoming the barriers that prevent integration of these critical glucose data into day-to-day management remains a challenge.

Citing Articles

The benefits and accuracy of real-time continuous glucose monitoring in children and adolescents with type 1 diabetes attending a summer camp.

Kondo T, Senokuchi T, Morinaga J, Miyashita A, Yano M, Takeda H J Diabetes Investig. 2024; 16(1):154-162.

PMID: 39474860 PMC: 11693605. DOI: 10.1111/jdi.14337.


Evaluating HbA-to-average glucose conversion with patient-specific kinetic models for diverse populations.

Sato Imuro S, Sabharwal A, Bevier W, Kerr D Sci Rep. 2024; 14(1):22098.

PMID: 39333162 PMC: 11437029. DOI: 10.1038/s41598-024-72837-7.


The association of continuous glucose monitoring with glycemic parameters in patients with uncontrolled type 2 diabetes: A prospective observational study.

Chaubey A, Chaubey D, Dwivedi A, Dwivedi S, Mishra T J Family Med Prim Care. 2024; 13(8):3038-3041.

PMID: 39228534 PMC: 11368378. DOI: 10.4103/jfmpc.jfmpc_1950_23.


ROUTE-T1D: A behavioral intervention to promote optimal continuous glucose monitor use among racially minoritized youth with type 1 diabetes: Design and development.

Straton E, Bryant B, Kang L, Wang C, Barber J, Perkins A Contemp Clin Trials. 2024; 140:107493.

PMID: 38460913 PMC: 11065587. DOI: 10.1016/j.cct.2024.107493.


Utility and precision evidence of technology in the treatment of type 1 diabetes: a systematic review.

Jacobsen L, Sherr J, Considine E, Chen A, Peeling S, Hulsmans M Commun Med (Lond). 2023; 3(1):132.

PMID: 37794113 PMC: 10550996. DOI: 10.1038/s43856-023-00358-x.


References
1.
. Validation of measures of satisfaction with and impact of continuous and conventional glucose monitoring. Diabetes Technol Ther. 2010; 12(9):679-84. PMC: 3045572. DOI: 10.1089/dia.2010.0015. View

2.
Tamborlane W, Beck R, Bode B, Buckingham B, Peter Chase H, Clemons R . Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008; 359(14):1464-76. DOI: 10.1056/NEJMoa0805017. View

3.
Haugstvedt A, Wentzel-Larsen T, Graue M, Sovik O, Rokne B . Fear of hypoglycaemia in mothers and fathers of children with Type 1 diabetes is associated with poor glycaemic control and parental emotional distress: a population-based study. Diabet Med. 2010; 27(1):72-8. DOI: 10.1111/j.1464-5491.2009.02867.x. View

4.
Clarke W, Snyder A, Cox D . Maternal fear of hypoglycemia in their children with insulin dependent diabetes mellitus. J Pediatr Endocrinol Metab. 1998; 11 Suppl 1:189-94. DOI: 10.1515/jpem.1998.11.s1.189. View

5.
Beck R, Buckingham B, Miller K, Wolpert H, Xing D, Block J . Factors predictive of use and of benefit from continuous glucose monitoring in type 1 diabetes. Diabetes Care. 2009; 32(11):1947-53. PMC: 2768196. DOI: 10.2337/dc09-0889. View