Abstract

Aim: To assess the time to obtain reliable oxygen saturation readings by different pulse oximeters during neonatal resuscitation in the delivery room or NICU.

Methods: Prospective study comparing three different pulse oximeters: Masimo Radical-7 compared simultaneously with Ohmeda Biox 3700 or with Nellcor N395, in newborn infants who required resuscitation. Members of the research team placed the sensors for each of the pulse oximeters being compared simultaneously, one sensor on each foot of the same baby. Care provided routinely, without interference by the research team. The time elapsed until a reliable SpO2 was obtained was recorded using a digital chronometer. Statistical comparisons included chi-square and student's T-test.

Results: Thirty-two infants were enrolled; median gestational age 32 weeks. Seventeen paired measurements were made with the Radical-7 and Biox 3700; mean time to a stable reading was 20.2±7 sec for the Radical-7 and 74.2±12 sec for the Biox 3700 (p=0.02). The Radical-7 and the N- 395 were paired on 15 infants; the times to obtain a stable reading were 20.9±4 sec and 67.3±12 sec, respectively (p=0.03).

Conclusion: The time to a reliable reading obtained simultaneously in neonatal critical situations differs by the type of the pulse oximeter used, being significantly faster with Masimo Signal Extraction Technology. This may permit for better adjustments of inspired oxygen, aiding in the prevention of damage caused by unnecessary exposure to high or low oxygen.

Citing Articles

Use of Pulse Oximetry during Resuscitation of 230 Newborns-A Video Analysis.

Kolstad V, Pike H, Eilevstjonn J, Buskov F, Ersdal H, Rettedal S Children (Basel). 2023; 10(7).

PMID: 37508621 PMC: 10377843. DOI: 10.3390/children10071124.


Dyspnea: perfusion index and triage status.

Tulay C, Gurmen E J Clin Monit Comput. 2023; 37(4):1103-1108.

PMID: 37004661 PMC: 10066968. DOI: 10.1007/s10877-023-00995-6.


Introducing pulse oximetry for outpatient management of childhood pneumonia: An implementation research adopting a district implementation model in selected rural facilities in Bangladesh.

Rahman A, Ameen S, Hossain A, Perkins J, Jabeen S, Majid T EClinicalMedicine. 2022; 50:101511.

PMID: 35795715 PMC: 9251564. DOI: 10.1016/j.eclinm.2022.101511.


Success and time implications of SpO measurement through pulse oximetry among hospitalised children in rural Bangladesh: Variability by various device-, provider- and patient-related factors.

Rahman A, Ameen S, Hossain A, Jabeen S, Majid T, Afm A J Glob Health. 2022; 12:04036.

PMID: 35493782 PMC: 9041243. DOI: 10.7189/jogh.12.04036.


In-Ear SpO: A Tool for Wearable, Unobtrusive Monitoring of Core Blood Oxygen Saturation.

Davies H, Williams I, Peters N, Mandic D Sensors (Basel). 2020; 20(17).

PMID: 32872310 PMC: 7506719. DOI: 10.3390/s20174879.


References
1.
Mariani G, Brener Dik P, Ezquer A, Aguirre A, Esteban M, Perez C . Pre-ductal and post-ductal O2 saturation in healthy term neonates after birth. J Pediatr. 2007; 150(4):418-21. DOI: 10.1016/j.jpeds.2006.12.015. View

2.
Sola A . Oxygen in neonatal anesthesia: friend or foe?. Curr Opin Anaesthesiol. 2008; 21(3):332-9. DOI: 10.1097/ACO.0b013e3282f8ad8d. View

3.
Malviya S, Reynolds P, Voepel-Lewis T, Siewert M, WATSON D, Tait A . False alarms and sensitivity of conventional pulse oximetry versus the Masimo SET technology in the pediatric postanesthesia care unit. Anesth Analg. 2000; 90(6):1336-40. DOI: 10.1097/00000539-200006000-00013. View

4.
Sola A, Saldeno Y, Favareto V . Clinical practices in neonatal oxygenation: where have we failed? What can we do?. J Perinatol. 2008; 28 Suppl 1:S28-34. DOI: 10.1038/jp.2008.47. View

5.
Wang C, Anderson C, Leone T, Rich W, Govindaswami B, Finer N . Resuscitation of preterm neonates by using room air or 100% oxygen. Pediatrics. 2008; 121(6):1083-9. DOI: 10.1542/peds.2007-1460. View