Accuracy of Pulse Oximetry Measurement of Heart Rate of Newborn Infants in the Delivery Room

Overview
Journal J Pediatr
Specialty Pediatrics
Date 2008 May 22
PMID 18492509
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To determine the accuracy of heart rate obtained by pulse oximetry (HR(PO)) relative to HR obtained by 3-lead electrocardiography (HR(ECG)) in newborn infants in the delivery room.

Study Design: Immediately after birth, a preductal PO sensor and ECG leads were applied. PO and ECG monitor displays were recorded by a video camera. Two investigators reviewed the videos. Every two seconds, 1 of the investigators recorded HR(PO) and indicators of signal quality from the oximeter while masked to ECG, whereas the other recorded HR(ECG) and ECG signal quality while masked to PO. HR(PO) and HR(ECG) measurements were compared using Bland-Altman analysis.

Results: We attended 92 deliveries; 37 infants were excluded due to equipment malfunction. The 55 infants studied had a mean (+/-standard deviation [SD]) gestational age of 35 (+/-3.7) weeks, and birth weight 2399 (+/-869) g. In total, we analyzed 5877 data pairs. The mean difference (+/-2 SD) between HR(ECG) and HR(PO) was -2 (+/-26) beats per minute (bpm) overall and -0.5 (+/-16) bpm in those infants who received positive-pressure ventilation and/or cardiac massage. The sensitivity and specificity of PO for detecting HR(ECG) <100 bpm was 89% and 99%, respectively.

Conclusion: PO provided an accurate display of newborn infants' HR in the delivery room, including those infants receiving advanced resuscitation.

Citing Articles

Does early heart rate detection and continuous monitoring have an impact on neonatal resuscitation in newborns with intact cord? - An observational study.

Choudhary S, Singh A, Pandey A, Gupta N, Kumar A, Kabisatpathy S J Family Med Prim Care. 2025; 13(12):5655-5661.

PMID: 39790764 PMC: 11709070. DOI: 10.4103/jfmpc.jfmpc_752_24.


Delivery Room Handling of the Newborn: Filling the Gaps.

Saugstad O, Kapadia V, Vento M Neonatology. 2024; 121(5):553-561.

PMID: 39308394 PMC: 11446302. DOI: 10.1159/000540079.


Fast and accurate newborn heart rate monitoring at birth: A systematic review.

Kapadia V, Kawakami M, Strand M, Hurst C, Spencer A, Schmolzer G Resusc Plus. 2024; 19:100668.

PMID: 38912532 PMC: 11190559. DOI: 10.1016/j.resplu.2024.100668.


2023 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations: Summary From the Basic Life Support; Advanced Life Support; Pediatric Life Support; Neonatal Life Support;....

Berg K, Bray J, Ng K, Liley H, Greif R, Carlson J Circulation. 2023; 148(24):e187-e280.

PMID: 37942682 PMC: 10713008. DOI: 10.1161/CIR.0000000000001179.


Significance of Neonatal Heart Rate in the Delivery Room-A Review.

Nerdrum Aagaard E, Solevag A, Saugstad O Children (Basel). 2023; 10(9).

PMID: 37761512 PMC: 10528538. DOI: 10.3390/children10091551.