Introducing Pulse Oximetry for Outpatient Management of Childhood Pneumonia: An Implementation Research Adopting a District Implementation Model in Selected Rural Facilities in Bangladesh
Overview
Authors
Affiliations
Background: Pulse oximetry has potential for identifying hypoxaemic pneumonia and substantially reducing under-five deaths in low- and middle-income countries (LMICs) setting. However, there are few examples of introducing pulse oximetry in resource-constrained paediatric outpatient settings, such as Integrated Management of Childhood Illness (IMCI) services.
Methods: The National IMCI-programme of Bangladesh designed and developed a district implementation model for introducing pulse oximetry in routine IMCI services through stakeholder engagement and demonstrated the model in Kushtia district adopting a health system strengthening approach. Between December 2020 and June 2021, two rounds of assessment were conducted based on WHO's implementation research framework and outcome variables, involving 22 IMCI service-providers and 1680 children presenting with cough/difficulty-in-breathing in 12 health facilities. The data collection procedures included structured-observations, re-assessments, interviews, and data-extraction by trained study personnel.
Findings: We observed that IMCI service-providers conducted pulse oximetry assessments on all eligible children in routine outpatient settings, of which 99% of assessments were successful; 85% (95% CI 83,87) in one attempt, and 69% (95% CI 67,71) within one minute. The adherence to standard operating procedure related to pulse oximetry was 92% (95% CI 91,93), and agreement regarding identifying hypoxaemia was 97% (95% CI 96,98). The median performance-time was 36 seconds (IQR 20,75), which was longer among younger children (2-11 months: 44s, IQR 22,78; 12-59 months: 30s, IQR 18,53, < 0.01) and among those classified as pneumonia/severe-pneumonia than as no-pneumonia (41s, IQR 22,70; 32s, IQR 20,62, < 0.01). We observed improvements in almost all indicators in round-2. IMCI service-providers and caregivers showed positive attitudes towards using this novel technology for assessing their children.
Interpretation: This implementation research study suggested the adoption, feasibility, fidelity, appropriateness, acceptability, and sustainability of pulse oximetry introduction in routine IMCI services in resource-poor settings. The learning may inform the evidence-based scale-up of pulse oximetry linked with an oxygen delivery system in Bangladesh and other LMICs.
Funding: This research was funded by the UK National Institute for Health Research (NIHR) (Global Health Research Unit on Respiratory Health (RESPIRE); 16/136/109) using UK aid from the UK Government to support global health research.
Graham H, King C, Rahman A, Kitutu F, Greenslade L, Aqeel M Lancet Glob Health. 2025; 13(3):e528-e584.
PMID: 39978385 PMC: 11865010. DOI: 10.1016/S2214-109X(24)00496-0.
Bauler S, Nivyindika L, Kirwa T, Habonimana V, Nizigiyimana D, Kirby M PLOS Glob Public Health. 2025; 5(1):e0002399.
PMID: 39804948 PMC: 11729966. DOI: 10.1371/journal.pgph.0002399.
Graham H, King C, Duke T, Ahmed S, Baqui A, Colbourn T Lancet Glob Health. 2024; 12(8):e1359-e1364.
PMID: 38914087 PMC: 11254785. DOI: 10.1016/S2214-109X(24)00209-2.
Kortz T, Mediratta R, Smith A, Nielsen K, Agulnik A, Rivera S Front Pediatr. 2024; 12:1397232.
PMID: 38910960 PMC: 11190367. DOI: 10.3389/fped.2024.1397232.
Jabeen S, Rahman M, Siddique A, Hasan M, Matin R, Rahman Q J Glob Health. 2024; 14:04075.
PMID: 38722093 PMC: 11082830. DOI: 10.7189/jogh.14.04075.