» Articles » PMID: 20824407

The Incidence of Acetabular Osteolysis in Young Patients with Conventional Versus Highly Crosslinked Polyethylene

Overview
Publisher Wolters Kluwer
Specialty Orthopedics
Date 2010 Sep 9
PMID 20824407
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Osteolysis is a major mode of hip implant failure. Previous literature has focused on the amount of polyethylene wear comparing highly crosslinked polyethylene (HXPLE) with conventional liners but has not clarified the relative incidence of osteolysis with these two liners.

Questions/purposes: We determined (1) the incidence of osteolysis in HXLPE versus conventional polyethylene (CPE), (2) the ability to detect and evaluate the size of lytic lesions using radiographs compared with CT scans, (3) head penetration in hips without and with lysis, and (4) determined whether acetabular position, head size, and UCLA activity score contributed to lysis.

Methods: We compared head penetration and osteolysis on plain radiographs and presence and volume of osteolysis on CT scans in 48 patients with HXLPE (mean, 46.5 years) and 50 patients with CPE (mean, 43.2 years). The minimum followup was 5 years (average, 7.2 years; range, 5.1-10.9 years),

Results: Osteolysis was apparent on CT in a larger number of patients with CPE liners than HXLPE liners: 12 of 50 (24%) versus one of 48 (2%), respectively. We found no correlation between head penetration and volume of osteolytic lesions. Head penetration was greater in patients with osteolysis. Smaller head sizes were associated with greater wear and those with osteolysis had smaller head sizes; however, there was no difference in acetabular component position or UCLA activity in those with lysis compared with those without.

Conclusions: HXLPE diminished the incidence of osteolysis, but the lack of correlation between penetration and volume of osteolysis suggests other factors other than wear contribute to the development of osteolysis.

Level Of Evidence: Level III, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

Citing Articles

Isolated Liner Exchange and Bone Grafting for the Management of Periacetabular Osteolysis in Well-Fixed Cups with an Intact Locking Mechanism at Short-Term to Medium-Term Follow-Up: A Systematic Review.

Ricotti R, Alexander-Malahias M, Ma Q, Jang S, Loucas R, Gkiatas I HSS J. 2024; 20(4):567-576.

PMID: 39494435 PMC: 11528827. DOI: 10.1177/15563316231189736.


[Long-term computed tomographic osteolytic analysis of highly cross-linked polyethylene prosthesis after total hip arthroplasty].

Wang X, Liu J Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2024; 38(9):1079-1085.

PMID: 39300882 PMC: 11440164. DOI: 10.7507/1002-1892.202403040.


Rate of revision and wear penetration in different polyethylene liner compositions in total hip arthroplasty: a Bayesian network meta-analysis.

Migliorini F, Betsch M, Maffulli N, Schafer L, Hildebrand F, Kubach J Sci Rep. 2024; 14(1):21162.

PMID: 39256531 PMC: 11387631. DOI: 10.1038/s41598-024-71326-1.


Mid-term outcomes of total hip arthroplasty in patients younger than 30 years.

Katzman J, Manjunath A, Brandstetter A, Ben-Ari E, Schwarzkopf R, Snir N Arch Orthop Trauma Surg. 2024; 144(8):3775-3786.

PMID: 39096326 DOI: 10.1007/s00402-024-05473-5.


Investigation of contact behavior on a model of the dual-mobility artificial hip joint for Asians in different inner liner thicknesses.

Hidayat T, Ammarullah M, Ismail R, Saputra E, Lamura M, K N C World J Orthop. 2024; 15(4):321-336.

PMID: 38680676 PMC: 11045469. DOI: 10.5312/wjo.v15.i4.321.


References
1.
Engh Jr C, Stepniewski A, Ginn S, Beykirch S, Sychterz-Terefenko C, Hopper Jr R . A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006; 21(6 Suppl 2):17-25. DOI: 10.1016/j.arth.2006.05.002. View

2.
Kitamura N, Naudie D, Leung S, Hopper Jr R, Engh Sr C . Diagnostic features of pelvic osteolysis on computed tomography: the importance of communication pathways. J Bone Joint Surg Am. 2005; 87(7):1542-50. DOI: 10.2106/JBJS.D.02882. View

3.
Puri L, Wixson R, Stern S, Kohli J, Hendrix R, Stulberg S . Use of helical computed tomography for the assessment of acetabular osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 2002; 84(4):609-14. DOI: 10.2106/00004623-200204000-00016. View

4.
Shon W, Gupta S, Biswal S, Han S, Hong S, Moon J . Pelvic osteolysis relationship to radiographs and polyethylene wear. J Arthroplasty. 2008; 24(5):743-50. DOI: 10.1016/j.arth.2008.02.012. View

5.
DAntonio J, Manley M, Capello W, Bierbaum B, Ramakrishnan R, Naughton M . Five-year experience with Crossfire highly cross-linked polyethylene. Clin Orthop Relat Res. 2005; 441:143-50. DOI: 10.1097/00003086-200512000-00024. View