A Robust Uniaxial Force Sensor for Minimally Invasive Surgery

Overview
Date 2010 Feb 23
PMID 20172798
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

This paper presents a novel miniature uniaxial force sensor for use within a beating heart during mitral valve annuloplasty. The sensor measures 5.5 mm in diameter and 12 mm in length and provides a hollow core to pass instrumentation. A soft elastomer flexure design maintains a waterproof seal. Fiber optic transduction eliminates electrical circuitry within the heart, and acetal components minimize ultrasound-imaging artifacts. Calibration uses a nonlinear viscoelastic method, and in vitro tests demonstrate a 0-4-N force range with rms errors of 0.13 N (< 3.2%). In vivo tests provide the first endocardial measurements of tissue-minimally invasive surgery instrument interaction forces in a beating heart.

Citing Articles

Recent Advances in Biomedical Photonic Sensors: A Focus on Optical-Fibre-Based Sensing.

Ochoa M, Algorri J, Roldan-Varona P, Rodriguez-Cobo L, Lopez-Higuera J Sensors (Basel). 2021; 21(19).

PMID: 34640788 PMC: 8513032. DOI: 10.3390/s21196469.


Optical Fiber Array Sensor for Force Estimation and Localization in TAVI Procedure: Design, Modeling, Analysis and Validation.

Bandari N, Dargahi J, Packirisamy M Sensors (Basel). 2021; 21(16).

PMID: 34450813 PMC: 8398250. DOI: 10.3390/s21165377.


Tool-tissue forces in surgery: A systematic review.

Golahmadi A, Khan D, Mylonas G, Marcus H Ann Med Surg (Lond). 2021; 65:102268.

PMID: 33898035 PMC: 8058906. DOI: 10.1016/j.amsu.2021.102268.


Tactile Perception Technologies and Their Applications in Minimally Invasive Surgery: A Review.

Huang C, Wang Q, Zhao M, Chen C, Pan S, Yuan M Front Physiol. 2021; 11:611596.

PMID: 33424634 PMC: 7785975. DOI: 10.3389/fphys.2020.611596.


A Piezoelectric Tactile Sensor for Tissue Stiffness Detection with Arbitrary Contact Angle.

Zhang Y, Ju F, Wei X, Wang D, Wang Y Sensors (Basel). 2020; 20(22).

PMID: 33218118 PMC: 7698970. DOI: 10.3390/s20226607.


References
1.
Reade C, Bower C, Bailey B, Maziarz D, Masroor S, Kypson A . Robotic mitral valve annuloplasty with double-arm nitinol U-clips. Ann Thorac Surg. 2005; 79(4):1372-6. DOI: 10.1016/j.athoracsur.2004.02.072. View

2.
Valdastri P, Harada K, Menciassi A, Beccai L, Stefanini C, Fujie M . Integration of a miniaturised triaxial force sensor in a minimally invasive surgical tool. IEEE Trans Biomed Eng. 2006; 53(11):2397-400. DOI: 10.1109/TBME.2006.883618. View

3.
Novotny P, Stoll J, Vasilyev N, Del Nido P, Dupont P, Zickler T . GPU based real-time instrument tracking with three-dimensional ultrasound. Med Image Anal. 2007; 11(5):458-64. PMC: 2693901. DOI: 10.1016/j.media.2007.06.009. View

4.
Yuen S, Kettler D, Novotny P, Plowes R, Howe R . Robotic Motion Compensation for Beating Heart Intracardiac Surgery. Int J Rob Res. 2010; 28(10):1355-1372. PMC: 2860971. DOI: 10.1177/0278364909104065. View