Benefit and Accuracy of Intraoperative 3D-imaging After Pedicle Screw Placement: a Prospective Study in Stabilizing Thoracolumbar Fractures

Overview
Journal Eur Spine J
Specialty Orthopedics
Date 2009 Jun 11
PMID 19513764
Citations 23
Authors
Affiliations
Soon will be listed here.
Abstract

Internal fixation is the established dorsal standard procedure for the treatment of thoracolumbar fractures. The main problem of the procedure is the false positioning of the pedicle screws. The exact determination of pedicle screws has up to now only been possible through postoperative computed tomography. This study was intended to clarify the diagnostic value of intraoperative 3D scans after pedicle screw implantation in thoracolumbar spine surgery. The direct intraoperative consequences of the 3D scans are reported and the results of the 3D scans are compared with the postoperative computed tomography images. Intraoperative 3D scans were prospectively carried out from June 2006 to October 2008 on 95 patients with fractures of the thoracolumbar spine that have been treated with internal fixation. Screws positions were categorised intraoperatively, screws in relevant malposition were repositioned immediately. A computed tomography of the involved spinal section was carried out postoperatively for all patients. The positions of the pedicle screws were determined and compared in the axial reconstructions of both procedures. Four hundred and fourteen pedicles with enclosed screws were evaluated by the 3D scans. The time needed for carrying out the 3D scan amounts to an average of 8.2 min. Eleven screws (2.7%) in ten patients were primarily intraoperatively repositioned on the basis of the 3D scan evaluation. Two of 95 patients had to have false positions of the screws revised secondarily following evaluation of the computed tomographies. The secondary postoperative revision rate of the patients amounts to 2.1%. In relation to the number of screws, this is a revision rate of 0.5%. The postoperative computed tomographies showed 323 pedicles without cortical penetration by the screws (78.0%). Ninety-one screws penetrated the pedicle wall (22%). It was possible to postoperatively compare the position classifications of 406 pedicle screws. The CT showed 378 correct screw positions, while 28 screws were positioned falsely. On the basis of the 3D scans, 376 of 378 correct positions were correctly assessed. Twenty-one of 28 false positions could be correctly classified. The sensitivity of all 3D scans reached 91.3% and the specificity 98.2%. The position of 97.8% of the pedicle screws was correctly recognised by the intraoperative 3D scan. Nine screws were classified falsely (2.2%). The comparison of the classification results showed significantly higher error findings by the 3D scan in the spinal section T1-10 (P = 0.014). The image quality of the 3D scan correlates significantly with the width of the scanned pedicle, with the body mass index, the scanned spinal section and the extent of the fixation assembly. 3D scans showed a high accuracy in predicting pedicle screw position. Primary false placement of screws and primary neurovascular damage cannot be avoided. But intraoperative evaluation of the 3D scans resulted in a primary revision rate of 2.7% of the pedicle screws and we could lower the secondary revision rate to 0.5%.

Citing Articles

The Impact of Navigation in Lumbar Spine Surgery: A Study of Historical Aspects, Current Techniques and Future Directions.

Heydar A, Tanaka M, Prabhu S, Komatsubara T, Arataki S, Yashiro S J Clin Med. 2024; 13(16).

PMID: 39200805 PMC: 11354833. DOI: 10.3390/jcm13164663.


Navigation-Guided/Robot-Assisted Spinal Surgery: A Review Article.

Lee Y, Cho D, Kim K Neurospine. 2024; 21(1):8-17.

PMID: 38569627 PMC: 10992634. DOI: 10.14245/ns.2347184.592.


Advances and Evolving Challenges in Spinal Deformity Surgery.

Patel R, Yearley A, Isaac H, Chalif E, Chalif J, Zaidi H J Clin Med. 2023; 12(19).

PMID: 37835030 PMC: 10573859. DOI: 10.3390/jcm12196386.


X23D-Intraoperative 3D Lumbar Spine Shape Reconstruction Based on Sparse Multi-View X-ray Data.

Jecklin S, Jancik C, Farshad M, Furnstahl P, Esfandiari H J Imaging. 2022; 8(10).

PMID: 36286365 PMC: 9604813. DOI: 10.3390/jimaging8100271.


Accuracy of pedicle screw placement using neuronavigation based on intraoperative 3D rotational fluoroscopy in the thoracic and lumbar spine.

Conrads N, Grunz J, Huflage H, Luetkens K, Feldle P, Grunz K Arch Orthop Trauma Surg. 2022; 143(6):3007-3013.

PMID: 35794344 PMC: 10191982. DOI: 10.1007/s00402-022-04514-1.


References
1.
Lonstein J, Denis F, Perra J, Pinto M, Smith M, Winter R . Complications associated with pedicle screws. J Bone Joint Surg Am. 1999; 81(11):1519-28. DOI: 10.2106/00004623-199911000-00003. View

2.
Laine T, Lund T, Ylikoski M, Lohikoski J, Schlenzka D . Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patients. Eur Spine J. 2000; 9(3):235-40. PMC: 3611394. DOI: 10.1007/s005860000146. View

3.
Vaccaro A, Kim D, Brodke D, Harris M, Chapman J, Schildhauer T . Diagnosis and management of thoracolumbar spine fractures. Instr Course Lect. 2004; 53:359-73. View

4.
Laine T, Makitalo K, Schlenzka D, Tallroth K, Poussa M, ALHO A . Accuracy of pedicle screw insertion: a prospective CT study in 30 low back patients. Eur Spine J. 1997; 6(6):402-5. PMC: 3467718. DOI: 10.1007/BF01834068. View

5.
Baldauf J, Muller J, Fleck S, Hinz P, Chiriac A, Schroeder H . The value of intraoperative three dimensional fluoroscopy in anterior decompressive surgery of the cervical spine. Zentralbl Neurochir. 2008; 69(1):30-4. DOI: 10.1055/s-2007-992796. View