[Studies on Image Quality, High Contrast Resolution and Dose for the Axial Skeleton and Limbs with a New, Dedicated CT System (ISO-C-3 D)]
Overview
Radiology
Authors
Affiliations
Purpose: Evaluation of 3D-CT imaging of the axial skeleton and different joints of the lower and upper extremities with a new dedicated CT system (ISO-C-3D) based on a mobile isocentric C-arm image amplifier.
Material And Methods: 27 cadaveric specimes of different joints of the lower and upper extremities and of the spinal column were examined with 3D-CT imaging (ISO-C-3D). All images were evaluated by 3 radiologists for image quality using a semiquantitative score (score value 1: poor quality; score value 4: excellent quality). In addition, dose measurements and measurements of high contrast resolution were performed in comparison to conventional and low-dose spiral CT using a high contrast phantom (Catphan, Phantom Laboratories).
Results: Adequate image quality (mean score values 3 - 4) could be achieved with an applied dose comparable to low-dose CT in smaller joints such as wrist, elbow, ankle and knee. A remarkably inferior image quality resulted in imaging of the hip, lumbar and thoracic spine (mean score values 2 - 3) in spite of almost doubling the dose (dose increased by 85 percent). The image quality of shoulder examinations was insufficient (mean score value 1). Phantom studies showed a high-contrast resolution comparable to helical CT in the xy-axis (9 lp/cm).
Conclusion: Preliminary results show, that image quality of C-arm-based CT-imaging (ISO-C-3D) seems to be adequate in smaller joints. ISO-C-3D images of the hip and axial skeleton show a decreased image quality, which does not seem to be sufficient for diagnosing subtle fractures.
Keil H, Beisemann N, Swartman B, Schnetzke M, Vetter S, Grutzner P Eur J Trauma Emerg Surg. 2022; 49(1):373-381.
PMID: 36048181 PMC: 9925545. DOI: 10.1007/s00068-022-02083-x.
Souleiman F, Henkelmann R, Theopold J, Fakler J, Spiegl U, Hepp P J Orthop Surg Res. 2021; 16(1):295.
PMID: 33952284 PMC: 8097962. DOI: 10.1186/s13018-021-02424-3.
Privalov M, Mohr M, Swartman B, Beisemann N, Keil H, Franke J J Digit Imaging. 2020; 33(5):1136-1143.
PMID: 32632648 PMC: 7572971. DOI: 10.1007/s10278-020-00324-2.
Beisemann N, Keil H, Swartman B, Schnetzke M, Franke J, Grutzner P J Orthop Surg Res. 2019; 14(1):236.
PMID: 31340818 PMC: 6657091. DOI: 10.1186/s13018-019-1286-7.
Jin L, Guo J, Guo J, Yin Y, Hou Z, Zhang Y Biomed Res Int. 2016; 2016:5898161.
PMID: 27975055 PMC: 5126400. DOI: 10.1155/2016/5898161.