Comparison Between Flotrac-Vigileo and Bioreactance, a Totally Noninvasive Method for Cardiac Output Monitoring

Overview
Journal Crit Care
Specialty Critical Care
Date 2009 May 21
PMID 19454009
Citations 45
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: This study was designed to compare the clinical acceptability of two cardiac output (CO) monitoring systems: a pulse wave contour-based system (FloTrac-Vigileo) and a bioreactance-based system (NICOM), using continuous thermodilution (PAC-CCO) as a reference method.

Methods: Consecutive patients, requiring PAC-CCO monitoring following cardiac surgery, were also monitored by the two other devices. CO values obtained simultaneously by the three systems were recorded continuously on a minute-by-minute basis.

Results: Continuous recording was performed on 29 patients, providing 12,099 simultaneous measurements for each device (417 +/- 107 per patient). In stable conditions, correlations of NICOM and Vigileo with PAC-CCO were 0.77 and 0.69, respectively. The bias was -0.01 +/- 0.84 for NICOM and -0.01 +/- 0.81 for Vigileo (NS). NICOM relative error was less than 30% in 94% of the patients and less than 20% in 79% vs. 91% and 79% for the Vigileo, respectively (NS). The variability of measurements around the trend line (precision) was not different between the three methods: 8 +/- 3%, 8 +/- 4% and 8 +/- 3% for PAC-CCO, NICOM and Vigileo, respectively. CO changes were 7.2 minutes faster with Vigileo and 6.9 minutes faster with NICOM (P < 0.05 both systems vs. PAC-CCO, NS). Amplitude of changes was not significantly different than thermodilution. Finally, the sensitivity and specificity for predicting significant CO changes were 0.91 and 0.95 respectively for the NICOM and 0.86 and 0.92 respectively for the Vigileo.

Conclusion: This study showed that the NICOM and Vigileo devices have similar monitoring capabilities in post-operative cardiac surgery patients.

Citing Articles

Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients.

Khanna A, Garcia J, Saha A, Harris L, Baruch M, Martin R J Clin Monit Comput. 2023; 38(1):139-146.

PMID: 37458916 DOI: 10.1007/s10877-023-01059-5.


Non-invasive continuous cardiac output monitoring in infants with hypoxic ischaemic encephalopathy.

Garvey A, ONeill R, Livingstone V, Pavel A, Finn D, Boylan G J Perinatol. 2022; 42(12):1622-1629.

PMID: 36056257 PMC: 9712087. DOI: 10.1038/s41372-022-01495-2.


Diagnostic Accuracy of Ultrasonographic Respiratory Variation in the Inferior Vena Cava, Subclavian Vein, Internal Jugular Vein, and Femoral Vein Diameter to Predict Fluid Responsiveness: A Systematic Review and Meta-Analysis.

Kim D, Chung S, Kang W, Kim J Diagnostics (Basel). 2022; 12(1).

PMID: 35054215 PMC: 8774961. DOI: 10.3390/diagnostics12010049.


Bioreactance reliably detects preload responsiveness by the end-expiratory occlusion test when averaging and refresh times are shortened.

Gavelli F, Beurton A, Teboul J, De Vita N, Azzolina D, Shi R Ann Intensive Care. 2021; 11(1):133.

PMID: 34453633 PMC: 8401368. DOI: 10.1186/s13613-021-00920-7.


Agreement of Bioreactance Cardiac Output Monitoring With Thermodilution in Healthy Standing Horses.

Hopster K, Hurcombe S Front Vet Sci. 2021; 8:701339.

PMID: 34414227 PMC: 8369349. DOI: 10.3389/fvets.2021.701339.


References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Critchley L, Critchley J . A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 2003; 15(2):85-91. DOI: 10.1023/a:1009982611386. View

3.
Squara P . Matching total body oxygen consumption and delivery: a crucial objective?. Intensive Care Med. 2004; 30(12):2170-9. DOI: 10.1007/s00134-004-2449-4. View

4.
Haller M, Zollner C, Briegel J, Forst H . Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study. Crit Care Med. 1995; 23(5):860-6. DOI: 10.1097/00003246-199505000-00014. View

5.
Keren H, Burkhoff D, Squara P . Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol. 2007; 293(1):H583-9. DOI: 10.1152/ajpheart.00195.2007. View