Noninvasive Cardiac Output Monitoring (NICOM): a Clinical Validation

Overview
Specialty Critical Care
Date 2007 Apr 27
PMID 17458538
Citations 93
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate the clinical utility of a new device for continuous noninvasive cardiac output monitoring (NICOM) based on chest bio-reactance compared with cardiac output measured semi-continuously by thermodilution using a pulmonary artery catheter (PAC-CCO).

Design: Prospective, single-center study.

Setting: Intensive care unit.

Patients: Consecutive adult patients immediately after cardiac surgery.

Interventions: Cardiac output measurements obtained from NICOM and thermodilution were simultaneously recorded minute by minute and compared in 110 patients. We evaluated the accuracy, precision, responsiveness, and reliability of NICOM for detecting cardiac output changes. Tolerance for each of these parameters was specified prospectively.

Measurements And Results: A total of 65,888 pairs of cardiac output measurements were collected. Mean reference values for cardiac output ranged from 2.79 to 9.27 l/min. During periods of stable PAC-CCO (slope<+/-10%, 2SD/mean<20%), the correlation between NICOM and thermodilution was R=0.82; bias was +0.16+/-0.52 l/min (+4.0+/-11.3%), and relative error was 9.1%+/-7.8%. In 85% of patients the relative error was <20%. During periods of increasing output, slopes were similar with the two methods in 96% of patients and intra-class correlation was positive in 96%. Corresponding values during periods of decreasing output were 90% and 84%, respectively. Precision was always better with NICOM than with thermodilution. During hemodynamic challenges, changes were 3.1+/-3.8 min faster with NICOM (p<0.01) and amplitude of changes did not differ significantly. Finally, sensitivity of the NICOM for detecting significant directional changes was 93% and specificity was 93%.

Conclusion: Cardiac output measured by NICOM had most often acceptable accuracy, precision, and responsiveness in a wide range of circulatory situations.

Citing Articles

Non-invasive continuous cardiac output monitoring in thoracic cancer surgery: A comparative study between calibrated pulse contour analysis and chest bioreactance.

Fellahi J, Abraham P, Tiberghien N, Coelembier C, Maury J, Bendjelid K Eur J Anaesthesiol Intensive Care. 2025; 1(4):e006.

PMID: 39917369 PMC: 11783632. DOI: 10.1097/EA9.0000000000000006.


Contemporary Review of Hemodynamic Monitoring in the Critical Care Setting.

Rali A, Butcher A, Tedford R, Sinha S, Mekki P, Van Spall H US Cardiol. 2024; 16:e12.

PMID: 39600839 PMC: 11588176. DOI: 10.15420/usc.2021.34.


Dynamic monitoring tools for patients admitted to the emergency department with circulatory failure: narrative review with panel-based recommendations.

Douglas I, Elwan M, Najarro M, Romagnoli S Eur J Emerg Med. 2024; 31(2):98-107.

PMID: 38364037 PMC: 11232941. DOI: 10.1097/MEJ.0000000000001103.


Age, sex, endurance capacity, and chronic heart failure affect central and peripheral factors of oxygen uptake measured by non-invasive and continuous technologies: support of pioneer work using invasive or non-continuous measures.

Brochhagen J, Coll Barroso M, Baumgart C, Wasmus D, Freiwald J, Hoppe M Front Sports Act Living. 2023; 5:1218948.

PMID: 37731479 PMC: 10507902. DOI: 10.3389/fspor.2023.1218948.


The Effect of Pharmacologically Induced Blood Pressure Manipulation on Cardiac Output and Cerebral Blood Flow Velocity in Patients with Aneurysmal Subarachnoid Hemorrhage.

Lakshmegowda M, Muthuchellapan R, Sharma M, Ganne S, Chakrabarti D, Muthukalai S Indian J Crit Care Med. 2023; 27(4):254-259.

PMID: 37378034 PMC: 10291643. DOI: 10.5005/jp-journals-10071-24435.


References
1.
Bernstein D . Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedance. Crit Care Med. 1986; 14(10):898-901. DOI: 10.1097/00003246-198610000-00015. View

2.
Rubini A, Del Monte D, Catena V, Attar I, Cesaro M, Soranzo D . Cardiac output measurement by the thermodilution method: an in vitro test of accuracy of three commercially available automatic cardiac output computers. Intensive Care Med. 1995; 21(2):154-8. DOI: 10.1007/BF01726539. View

3.
Nelson L . The new pulmonary artery catheters: continuous venous oximetry, right ventricular ejection fraction, and continuous cardiac output. New Horiz. 1997; 5(3):251-8. View

4.
Rocco M, Spadetta G, Morelli A, DellUtri D, Porzi P, Conti G . A comparative evaluation of thermodilution and partial CO2 rebreathing techniques for cardiac output assessment in critically ill patients during assisted ventilation. Intensive Care Med. 2003; 30(1):82-7. DOI: 10.1007/s00134-003-2069-4. View

5.
Spiess B, Patel M, Soltow L, Wright I . Comparison of bioimpedance versus thermodilution cardiac output during cardiac surgery: evaluation of a second-generation bioimpedance device. J Cardiothorac Vasc Anesth. 2001; 15(5):567-73. DOI: 10.1053/jcan.2001.26533. View