3D Diffusion Tensor MRI with Isotropic Resolution Using a Steady-state Radial Acquisition

Overview
Date 2009 Apr 24
PMID 19388107
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To obtain diffusion tensor images (DTI) over a large image volume rapidly with 3D isotropic spatial resolution, minimal spatial distortions, and reduced motion artifacts, a diffusion-weighted steady-state 3D projection (SS 3DPR) pulse sequence was developed.

Materials And Methods: A diffusion gradient was inserted in a SS 3DPR pulse sequence. The acquisition was synchronized to the cardiac cycle, linear phase errors were corrected along the readout direction, and each projection was weighted by measures of consistency with other data. A new iterative parallel imaging reconstruction method was also implemented for removing off-resonance and undersampling artifacts simultaneously.

Results: The contrast and appearance of both the fractional anisotropy and eigenvector color maps were substantially improved after all correction techniques were applied. True 3D DTI datasets were obtained in vivo over the whole brain (240 mm field of view in all directions) with 1.87 mm isotropic spatial resolution, six diffusion encoding directions in under 19 minutes.

Conclusion: A true 3D DTI pulse sequence with high isotropic spatial resolution was developed for whole brain imaging in under 20 minutes. To minimize the effects of brain motion, a cardiac synchronized, multiecho, DW-SSFP pulse sequence was implemented. Motion artifacts were further reduced by a combination of linear phase correction, corrupt projection detection and rejection, sampling density reweighting, and parallel imaging reconstruction. The combination of these methods greatly improved the quality of 3D DTI in the brain.

Citing Articles

DeepSTI: Towards tensor reconstruction using fewer orientations in susceptibility tensor imaging.

Fang Z, Lai K, van Zijl P, Li X, Sulam J Med Image Anal. 2023; 87:102829.

PMID: 37146440 PMC: 10288385. DOI: 10.1016/j.media.2023.102829.


Diffusion-weighted double-echo steady-state with a three-dimensional cones trajectory for non-contrast-enhanced breast MRI.

Moran C, Cheng J, Sandino C, Carl M, Alley M, Rosenberg J J Magn Reson Imaging. 2020; 53(5):1594-1605.

PMID: 33382171 PMC: 8564805. DOI: 10.1002/jmri.27492.


Reducing slab boundary artifacts in three-dimensional multislab diffusion MRI using nonlinear inversion for slab profile encoding (NPEN).

Wu W, Koopmans P, Frost R, Miller K Magn Reson Med. 2015; 76(4):1183-95.

PMID: 26510172 PMC: 4854328. DOI: 10.1002/mrm.26027.


In vivo diffusion spectrum imaging of non-human primate brain: initial experience in transcallosal fiber examination.

Meng Y, Zhang X Quant Imaging Med Surg. 2014; 4(2):129-35.

PMID: 24834425 PMC: 4014874. DOI: 10.3978/j.issn.2223-4292.2014.04.05.


Real-time correction of rigid body motion-induced phase errors for diffusion-weighted steady-state free precession imaging.

OHalloran R, Aksoy M, Aboussouan E, Peterson E, Van A, Bammer R Magn Reson Med. 2014; 73(2):565-76.

PMID: 24715414 PMC: 4190128. DOI: 10.1002/mrm.25159.


References
1.
Gmitro A, Alexander A . Use of a projection reconstruction method to decrease motion sensitivity in diffusion-weighted MRI. Magn Reson Med. 1993; 29(6):835-8. DOI: 10.1002/mrm.1910290619. View

2.
Jung Y, Jashnani Y, Kijowski R, Block W . Consistent non-cartesian off-axis MRI quality: calibrating and removing multiple sources of demodulation phase errors. Magn Reson Med. 2006; 57(1):206-12. DOI: 10.1002/mrm.21092. View

3.
Liu J, Redmond M, Brodsky E, Alexander A, Lu A, Thornton F . Generation and visualization of four-dimensional MR angiography data using an undersampled 3-D projection trajectory. IEEE Trans Med Imaging. 2006; 25(2):148-57. DOI: 10.1109/TMI.2005.861706. View

4.
Enzmann D, Pelc N . Brain motion: measurement with phase-contrast MR imaging. Radiology. 1992; 185(3):653-60. DOI: 10.1148/radiology.185.3.1438741. View

5.
Jung K, Cho Z . Reduction of flow artifacts in NMR diffusion imaging using view-angle tilted line-integral projection reconstruction. Magn Reson Med. 1991; 19(2):349-60. DOI: 10.1002/mrm.1910190226. View