Distal Femoral Fixation: a Biomechanical Comparison of Trigen Retrograde Intramedullary (i.m.) Nail, Dynamic Condylar Screw (DCS), and Locking Compression Plate (LCP) Condylar Plate

Overview
Journal J Trauma
Specialty Emergency Medicine
Date 2009 Feb 11
PMID 19204519
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The purpose of this study was to establish if there are biomechanical differences between implants in stiffness of construct, microdisplacement, and fatigue failure in a supracondylar femoral fracture model.

Methods: A retrograde intramedullary (i.m.) nail, dynamic condylar screw (DCS), and locked condylar plate (LCP) were tested using 33-cm long synthetic femurs. A standardized supracondylar medial segmental defect was created in the distal femur bone models. A gap away from the distal joint axis and parallel to the knee axis was created for axial testing of the specimens (Arbeitsgemeinschaft fur Osteosynthesefragen [AO] type 33-A) and a T-fracture (33-C) was created for the fatigue testing of the specimens. Peak displacements were measured, and analysis was done to determine construct stiffness and gap micromotion in axial loading. Cyclic loading was performed for fatigue testing.

Results: It was observed that there were statistically significant differences in micromotion across the fracture gap and overall stiffness of various implant constructs. The stiffness of the i.m. nail, DCS, and LCP were 1,106, 750, and 625 N/mm, respectively. The average total micromotion across the fracture gap for the i.m. nail, DCS, and LCP were 1.96, 10.55, and 17.74 mm, respectively. In fatigue testing, the i.m. nail distal screws failed at 9,000 cycles, the DCS did not fail (80,000 cycles completed), and the LCP failed at 19,000 and 23,500 cycles.

Conclusions: When considering micromotion and construct stiffness, the i.m. nail had statistically significant higher stiffness and significantly lower micromotion across the fracture gap with axial compression. Hence, the i.m. nail tested had the greatest stability for type 33-A fractures. However, the nail demonstrated the least amount of resistance to fatigue failure with type 33-C fractures, whereas the DCS did not fail with testing in any pattern.

Citing Articles

Long-term functional outcomes of BIFOLD osteosynthesis in distal femoral fractures with metaphyseal comminution: A retrospective analysis.

Desouza C, Antao N, Londhe S, Banka P J Orthop. 2024; 63:16-20.

PMID: 39524105 PMC: 11549979. DOI: 10.1016/j.jor.2024.10.033.


Locked plate constructs are not necessarily stiffer than nonlocked constructs: A biomechanical investigation of locked versus nonlocked diaphyseal fixation in a human cadaveric model of nonosteoporotic and osteoporotic distal femoral fractures.

Ricci W, Dvorzhinskiy A, Zheng Y, Walker Z, Anthony M, Holbrook J OTA Int. 2024; 7(2):e308.

PMID: 38549842 PMC: 10969512. DOI: 10.1097/OI9.0000000000000308.


Outcome Evaluation of Distal Femoral Fractures Following Surgical Management: A Retrospective Cohort Study.

Neumann-Langen M, Sontheimer V, Borchert G, Izadpanah K, Schmal H, Kubosch E J Pers Med. 2023; 13(2).

PMID: 36836584 PMC: 9960625. DOI: 10.3390/jpm13020350.


The geriatric distal femur fracture: nail, plate or both?.

Wilson J, Squires M, McHugh M, Ahn J, Perdue A, Hake M Eur J Orthop Surg Traumatol. 2022; 33(5):1485-1493.

PMID: 35895117 DOI: 10.1007/s00590-022-03337-5.


Locked Plating Versus Retrograde Intramedullary Nailing for Distal Femur Fractures: a Systematic Review and Meta-Analysis.

Neradi D, Sodavarapu P, Jindal K, Kumar D, Kumar V, Goni V Arch Bone Jt Surg. 2022; 10(2):141-152.

PMID: 35655740 PMC: 9117898. DOI: 10.22038/abjs.2021.53515.2656.