Aim: To create a new, automated method of evaluating the quality of optical coherence tomography (OCT) images and to compare its image quality discriminating ability with the quality assessment parameters signal to noise ratio (SNR) and signal strength (SS).

Methods: A new OCT image quality assessment parameter, quality index (QI), was created. OCT images (linear macular scan, peripapillary circular scan, and optic nerve head scan) were analysed using the latest StratusOCT system. SNR and SS were collected for each image. QI was calculated based on image histogram information using a software program of our own design. To evaluate the performance of these parameters, the results were compared with subjective three level grading (excellent, acceptable, and poor) performed by three OCT experts.

Results: 63 images of 21 subjects (seven each for normal, early/moderate, and advanced glaucoma) were enrolled in this study. Subjects were selected in a consecutive and retrospective fashion from our OCT imaging database. There were significant differences in SNR, SS, and QI between excellent and poor images (p = 0.04, p = 0.002, and p<0.001, respectively, Wilcoxon test) and between acceptable and poor images (p = 0.02, p<0.001, and p<0.001, respectively). Only QI showed significant difference between excellent and acceptable images (p = 0.001). Areas under the receiver operating characteristics (ROC) curve for discrimination of poor from excellent/acceptable images were 0.68 (SNR), 0.89 (IQP), and 0.99 (QI).

Conclusion: A quality index such as QI may permit automated objective and quantitative assessment of OCT image quality that performs similarly to an expert human observer.

Citing Articles

Development of next generation low-cost OCT towards improved point-of-care retinal imaging.

Price H, Song G, Wang W, OKane E, Lu K, Jelly E Biomed Opt Express. 2025; 16(2):748-759.

PMID: 39958865 PMC: 11828462. DOI: 10.1364/BOE.551625.


High Prevalence of Artifacts in Optical Coherence Tomography With Adequate Signal Strength.

Lin W, Coyner A, Amankwa C, Lucero A, Wollstein G, Schuman J Transl Vis Sci Technol. 2024; 13(8):43.

PMID: 39196579 PMC: 11364177. DOI: 10.1167/tvst.13.8.43.


In vivo identification of the retinal layer containing photopigments in OCT images through correlation with two-photon psychophysics.

Bartuzel M, Consejo A, Stremplewski P, Sylwestrzak M, Szkulmowski M, Gorczynska I Sci Rep. 2024; 14(1):15459.

PMID: 38965299 PMC: 11224378. DOI: 10.1038/s41598-024-65234-7.


An AS-OCT image dataset for deep learning-enabled segmentation and 3D reconstruction for keratitis.

Sun Y, Maimaiti N, Xu P, Jin P, Cai J, Qian G Sci Data. 2024; 11(1):627.

PMID: 38871784 PMC: 11176413. DOI: 10.1038/s41597-024-03464-0.


Longitudinal imaging of vitreal hyperreflective foci in mice with acute optic nerve damage using visible-light optical coherence tomography.

Fan W, Miller D, Chang S, Kweon J, Yeo W, Grannonico M Opt Lett. 2024; 49(8):1880-1883.

PMID: 38621029 PMC: 11217911. DOI: 10.1364/OL.512029.


References
1.
Guedes V, Schuman J, Hertzmark E, Wollstein G, Correnti A, Mancini R . Optical coherence tomography measurement of macular and nerve fiber layer thickness in normal and glaucomatous human eyes. Ophthalmology. 2003; 110(1):177-89. PMC: 1949047. DOI: 10.1016/s0161-6420(02)01564-6. View

2.
Drexler W, Sattmann H, Hermann B, Ko T, Stur M, Unterhuber A . Enhanced visualization of macular pathology with the use of ultrahigh-resolution optical coherence tomography. Arch Ophthalmol. 2003; 121(5):695-706. DOI: 10.1001/archopht.121.5.695. View

3.
Paunescu L, Schuman J, Price L, Stark P, Beaton S, Ishikawa H . Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using StratusOCT. Invest Ophthalmol Vis Sci. 2004; 45(6):1716-24. PMC: 1993821. DOI: 10.1167/iovs.03-0514. View

4.
Huang D, Swanson E, Lin C, Schuman J, Stinson W, Chang W . Optical coherence tomography. Science. 1991; 254(5035):1178-81. PMC: 4638169. DOI: 10.1126/science.1957169. View

5.
Hee M, Izatt J, Swanson E, Huang D, Schuman J, Lin C . Optical coherence tomography of the human retina. Arch Ophthalmol. 1995; 113(3):325-32. DOI: 10.1001/archopht.1995.01100030081025. View