» Articles » PMID: 16239620

Quantitative Upper Airway Imaging with Anatomic Optical Coherence Tomography

Overview
Specialty Critical Care
Date 2005 Oct 22
PMID 16239620
Citations 47
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Measurements of upper airway size and shape are important in investigating the pathophysiology of obstructive sleep apnea (OSA) and in devising, applying, and determining the effectiveness of treatment modalities. We describe an endoscopic optical technique (anatomic optical coherence tomography, aOCT) that provides quantitative real-time imaging of the internal anatomy of the human upper airway.

Methods: Validation studies were performed by comparing aOCT- and computed tomography (CT)-derived measurements of cross-sectional area (CSA) in (1) conduits in a wax phantom and (2) the velo-, oro-, and hypopharynx during wakefulness in five volunteers. aOCT scanning was performed during sleep in one subject with OSA.

Results: aOCT generated images of pharyngeal shape and measurements of CSA and internal dimensions that were comparable to radiographic CT images. The mean difference between aOCT- and CT-derived measurements of CSA in (1) the wax phantom was 2.1 mm(2) with limits of agreement (2 SD) from -13.2 to 17.4 mm(2) and intraclass correlation coefficient of 0.99 (p < 0.001) and (2) the pharyngeal airway was 14.1 mm(2) with limits of agreement from -43.7 to 57.8 mm(2) and intraclass correlation coefficient of 0.89 (p < 0.001). aOCT generated quantitative images of changes in upper airway size and shape before, during, and after an apneic event in an individual with OSA.

Conclusions: aOCT generates quantitative, real-time measurements of upper airway size and shape with minimal invasiveness, allowing study over lengthy periods during both sleep and wakefulness. These features should make it useful for study of upper airway behavior to investigate OSA pathophysiology and aid clinical management.

Citing Articles

A review on diagnostic assessments of tracheal stenosis.

Monjezi M, Rismanian M, Jamaati H Biomed Eng Online. 2025; 24(1):18.

PMID: 39953580 PMC: 11827378. DOI: 10.1186/s12938-025-01351-2.


A systematic review of the techniques for automatic segmentation of the human upper airway using volumetric images.

Maken P, Gupta A, Gupta M Med Biol Eng Comput. 2023; 61(8):1901-1927.

PMID: 37248380 DOI: 10.1007/s11517-023-02842-x.


Passively scanned, single-fiber optical coherence tomography probes for gastrointestinal devices.

Otuya D, Dechene N, Poshtupaka D, Judson S, Carlson C, Zemlok S Lasers Surg Med. 2022; 54(7):935-944.

PMID: 35708124 PMC: 9541095. DOI: 10.1002/lsm.23576.


Automatic correction of the initial rotation angle error improves 3D reconstruction in endoscopic airway optical coherence tomography.

Qi L, Zhuang Z, Zhang S, Huang S, Feng Q, Chen W Biomed Opt Express. 2022; 12(12):7616-7631.

PMID: 35003856 PMC: 8713659. DOI: 10.1364/BOE.439120.


Small airway dilation measured by endoscopic optical coherence tomography correlates with chronic lung allograft dysfunction.

Malone J, Lee A, Hohert G, Nador R, Lane P J Biomed Opt. 2021; 26(7).

PMID: 34263577 PMC: 8278781. DOI: 10.1117/1.JBO.26.7.076005.